Stopped-Flow Spectrophotometer System
Awarded Award Notice Posted
- Solicitation number
- RFQ-OH-17-00171
- Agency
- Cincinnati Procurement Operations Division Office of Acquisition Management, Environmental Protection Agency
- Awarded
- to Applied Photophysics LTD
- Set-aside
- No set-aside
Opportunity facts
- Contract number
- EP18C000008 Federal contract award
- NAICS code
- 334516 Analytical Laboratory Instrument Manufacturing
- PSC
- Not on record
Notice details come from SAM.gov. Updated .
Notice text
2 versions
Update #2 · Latest ·
Added: Jul 18, 2017 11:03 am Modified: Aug 31, 2017 10:49 amTrack Changes Stopped-Flow Spectrophotometer System
Update #1 ·
Added: Jul 18, 2017 11:03 am
Notice Type: Pre-solicitation
Synopsis: Stopped-Flow Spectrophotometer System
NAICS Code: 334516
Product Service Code: 6640
The U.S. Environmental Protection Agency (EPA), Office of Research and Development's (ORD), National Risk Management Research Laboratory (NRMRL), Water Systems Division (WSD) has a requirement for a Stopped-Flow Spectrophotometer System (SFSS) for performing in-house research. The SFSS is a fully automated system, which allows absorbance, fluorescence, or conductivity to be used to determine fast reaction kinetics.
The Stopped-Flow Spectrophotometer System must meet or exceed all of the following minimum specifications:
1. Single and double mixing stopped-flow
A fully automated, computer controlled stopped-flow system capable of absorbance and fluorescence measurements, single mixing, and double (sequential) mixing using three reagents and an age loop to incubate the first mixed solution for a set time (minimum aging time of 15 milliseconds or less and a maximum aging time of 1,000 seconds or greater) before reacting with a third reagent in the observation cell. The stopped-flow must be capable of a dead time of 1.2 milliseconds (ms) or less with a standard cell and 0.5 ms or less with a low volume cell.
-Must provide 3-4 syringes for sequential mixing.
-Must allow for asymmetric mixing, with ratio mixing up to at least 10:1.
-Must provide standard cell with dead time of 1.2 ms or less.
-Must provide cuvette geometry able to accommodate multiple detectors.
-Must provide pneumatic syringe system for fluid delivery.
-Must operate in a working temperature ranging from 0°C to 60°C.
-Must provide connectors compatible with water bath systems.
-Low volume cell for fast kinetics, up to 2000 s-1
-Must provide easily changeable low volume cell for fast kinetics with a dead time of 0.5 ms or less.
3. Multi-wavelength detector
-Must provide multi-wavelength detector with variable integration time for homogenous spectral data from a single shot.
-Must time resolve absorption spectra.
-Must operate for a wavelength range of at least 225-725 nm.
-Must allow a variable integration time.
4. Photomultiplier detection
-Must provide photomultiplier detection for fluorescence and for optimum sensitivity and speed in absorbance mode from 200¿850 nm.
5. Light range 200-850 nm
-Must provide light sources and stabilized power supply to cover the range of 200-850 nm.
6. Conductivity detector
-Must provide single and double mixing conductivity detection and be easily interchangeable.
-Must provide a dead time of 10 ms or less.
-Must cover a minimum conductivity range of 0-2 mS.
7. Temperature range from 0°C to 60°C
-Must work from 0°C to 60°C.
-Must provide provision to eliminate condensation at low temperatures.
8. Circulator and cooler
-Must include circulator and cooler to thermostat the system at the desired temperature from 0°C to 60°C.
9. Quench-Flow attachment
-Must include an attachment that allows the conversion of the sequential mixing system to a simple quench-flow unit where the user can recover aged, quenched samples.
-Must allow user-selected sample age times from 30 ms to 1,000 seconds.
10. Ratio mixing up to 10:1
-In addition to 1:1 mixing, the system must be capable of asymmetric mixing in single mixing mode, allowing ratio mixing up to 10:1.
-Must provide 2 x 0.5 mL syringes.
-Must provide 2 x 2.5 mL syringes.
11. Dual detection
-Must have the capability to obtain fluorescence and absorbance measurements simultaneously.
12. Fluorescence emission scanning
-Must have the capability to resolve fluorescence scanning emission measurements.
13. Instrument operating software and computer
-Must provide a high specification personal computer configured for system operation and with appropriate operating system software.
-Instrument Operating Software
o Must provide multiple licenses.
o Must allow data acquisition and analysis and full control of the stopped-flow system hardware.
o Must allow fitting of data using multiple models to estimate reaction rates.
o Must export data in Excel compatible files (.csv and .txt).
o Must be compatible with Windows 10.
14. Spares kit
-Must provide spares to cover consumables that may be necessary, including drive valves, range of fuses, syringes, and replacement lamps.
15. Installation and training
-Must provide 3 days of installation and on¿site training.
-Must provide hardware and software manuals in electronic and hard copy format.
16. Warranty
-Must provide 1 year warranty for any faulty parts that includes parts and labor starting from the date of instrument installation.
-Must provide on-site maintenance during the warranty period, if necessary, at 26 West Martin Luther King Drive, Cincinnati, Ohio 45268.
17. Shipping
-Must include shipping costs.
The government anticipates awarding a single fixed-price purchase order. This procurement will be accomplished under FAR Part 12 Commercial Procedures. THIS NOTICE OF INTENT IS NOT A REQUEST FOR COMPETITIVE QUOTATIONS. OFFERS WILL NOT BE ACCEPTED UNTIL THE SOLICITATION IS ISSUED THROUGH FEDCONNECT®. The solicitation is expected to be issued by the beginning of August 2017 through the FedConnect® portal. It is the responsibility of the offerors to frequently check the FedConnect® portal for updates or changes.
FedConnect® is a web portal that connects agencies and vendors to streamline the process of doing business with the federal government. Through FedConnect® contractors will be able to receive, review and respond to contract pre- and post-award actions and documents such as correspondence, request for proposals, tasking instruments and contract modifications. The use of FedConnect® also furthers the EPA's commitment to moving towards a paperless acquisition environment by reducing its carbon footprint and conducting its business in an ecologically friendly manner. There is no charge to use FedConnect®.
Vendors can register with FedConnect® at www.fedconnect.net/FedConnect/. For assistance in registering or for other FedConnect® technical questions please call the FedConnect® Help Desk at (800) 899-6665 or email at support@fedconnect.net.
Notice history
| Notice | Type | Posted |
|---|---|---|
| Stopped-Flow Spectrophotometer System | Award Notice | |
| Stopped-Flow Spectrophotometer System | Award Notice |
On GovTribe
Work this opportunity on GovTribe
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity