Multi-angle light scattering and viscometer detectors

Closed Solicitation Posted

Solicitation number
AG-3JL0-S-19-0010
Agency
Northeast Area Field Research Implementation and Information Delivery, Department of Agriculture
Responses due
Set-aside
No set-aside

Opportunity facts

NAICS code
334516 Analytical Laboratory Instrument Manufacturing
PSC
Not on record
Place of performance
Usda, ARS, Errc 600 East Mermaid Lane Wyndmoor, Pennsylvania 19038, United States

Notice details come from SAM.gov. Updated .

Notice text

66- MULTI-ANGLE LIGHT SCATTERING AND VISCOMETER DETECTORS
AG-3JL0-S-19-0010
08/21/2019
09/3/2019
Point of Contact, Regina Dennis, Contract Specialist, (215)233-6553 or e-mail regina.dennis@usda.gov
United States Department of Agriculture, Agricultural Research Service (ARS), Eastern Regional Research Center, Dairy and Functional Foods Research Unit (DFF) in Wyndmoor, PA 19038, is seeking to purchase a Multi-Angle Light Scattering and Viscometer Detectors to be furnished and delivered (F.O.B. Destination within Consignees Premises) to the USDA, ARS, Eastern Regional Research Center, in Wyndmoor, PA; in accordance with the terms, conditions, and specifications contained in this document.
This is a combined synopsis/solicitation for commercial items prepared in accordance with the format in FAR subpart 12.6 (current as of FAC 2019-05 eff. 8-13-2019), as supplemented with additional information included in this notice. This announcement constitutes the only solicitation quotations are being requested and a written solicitation will not be issued. The NAICS is 334516. The Solicitation number is AG-3JL0-S-19-0010 and is issued as a Request For Quotation (RFQ). SCHEDULE OF ITEMS - Furnish and deliver, F.O.B. Destination (shipping included).

Required Specifications for a Multi-Angle Light Scattering and Viscometer Detectors:
-Multi-angle Light Scattering (MALS) using detectors at 18 or more angles. It must be capable of determining the molar mass of polymers, oligosaccharides, proteins and peptides 500 Da-500 MDa. Must determine radius of gyration up to 400 nm.
-Viscometry - must limit baseline pressure drift to < 3 Pa/hr, Differential pressure baseline noise < 0. 1 Pa and provide temperature stability of.