MicroCT Scanner
Awarded Award Notice Posted
- Solicitation number
- VA26816R0021
- Awarded
- to Scanco USA Inc.
- Set-aside
- Total Small Business
Opportunity facts
- Contract number
- VA26816C0016 Federal contract award
- NAICS code
- 334516 Analytical Laboratory Instrument Manufacturing
- PSC
- Not on record
- Place of performance
- Philadelphia Vamc 19104 USA
Notice details come from SAM.gov. Updated .
Notice text
Added: Oct 22, 2015 2:58 pm
The Veteran's Administration (VA) Office of Research and Development, Program Contracting Office - East in Pittsburgh, PA intends to negotiate on a sole source basis with Scanco Medical USA in Wayne, PA. The requirement is to provide, ship, and install a Scanco microCT µCT 50 for objects up to 50 mm diameter specimen scanner at a VA facility in Philadelphia, PA. The sole source is pursuant to FAR 6.302-1 Only One Responsible Source and No Other Supplies or Services Will Satisfy Agency Requirements. This notice of intent is not a request for competitive proposals, however; any responsible source who believes it is capable of meeting the requirement may submit a capability statement and supporting documentation to the contracting office no later than Friday, Oct 30, 2015, (10:00 AM, EST). Interest/capability statements may be sent to Michael Haydo at Michael.haydo@va.gov. No telephone responses will be accepted. A determination not to compete the proposed requirement based upon the responses to this notice is solely within the discretion of the Government. The equipment and software described in the capabilities statement shall be comparable with the Scanco microCT µCT 50 system. The requirement characteristics are:
Equipment specifications:
" The system shall be capable of scanning large samples (up to 50mm in diameter - 120mm height) with nano-scale resolution.
" Detector system, data acquisition electronics (Physical Size 1700 x 1200; Virtual size: 3400 x 42000.
" Multi-position filter for x-ray source.
" A minimum of 30 total specimen holders to allow specimen scanning in liquid or air.
" A minimum of four (4) calibrated energy positions (ECP) between 40 kVp - 90 kVp.
" Server computer system controller with a minimum of 64 GB RAM, 3x 146 GB disks + external disk tower with a minimum of 40 TB data storage. A minimum of 1 x 10/100/1000Mb/s Ethernet. An open VMS operating system (Min 4-CPU license). Minimum 24" tft monitor. LTO6 Tape Drive.
" One (1) GPU Reconstruction Engine. Reconstruction software with license. Microsoft Windows workstation. 1 x K20 Kepler card1. Intel Xeon 2.7 GHz processors (or equal), Dual 12-core processor, 128 GB RAM, 1 TB storage.
" Software (64-bit)
" Measurement Program - Features including but not limited to:
a) Scout View with calibrated diameter positions
b) Single-Slice and Multi-Slice (3D) Scanning (interactive slice number selection)
c) High-throughput Batch Measurements (24 hour/7 Day unattended operation)
d) Single or Multiple Samples
" Fast Image Reconstruction:
a) On-line Reconstruction
b) Off-line Reconstruction
c) On-line beam-hardening correction
" Database Program:
a) User Names
b) Sample Names
c) Results (2D and 3D)
d) Back-up and Archive
" Image Analysis Software Tools - Features including but not limited to:
a) Zoom
b) Distance (2D and 3D)
c) Contouring, Morphing,
d) Automatic/Manual gray scale adjustment
e) Intensity based thresholding
f) Region based segmentation (geometric and/or conforming ROI's)
g) Cut/Paste Options
h) Morphing
i) Density Measurements (mg/cc HA, Hounsfield Units, Linear Attenuation Coefficient)
j) Masking
k) Histograms
" Morphometry Software:
a) Bone Area/Surface/Volume
b) Trabecular Thickness and Separation maps
c) Trabecular Number
d) Connectivity Density
e) Structural Model Index
f) Degree of Anisotropy
" Other functions:
a) Porosity
b) Bone-Implant Interface contact area/volume
c) Cortical Bone Thickness/Volume/Moment of Inertia calculations
d) Other functions provided on request
" 3D Image Rendering and Visualization:
a) Automatic or Manual 3D rendering
b) Animation
c) Transparency
d) Multiple color capability
" Image Output (Input to MATLAB/Analyze/Amira, and other 3rd party software) :
a) DICOM
b) Postscript
c) Tiff
d) JPEG
e) Other outputs (STL, etc.)
" Data Output:
a) Standard result sheets
b) Spreadsheet ready text files
" Networking:
a) Printing via TCPIP/DCPS/LPD/LPR protocols
b) Browser access to data, other functions
c) Remote access and control/use (via 3rd party Xwindows Software); unlimited software licenses for vendor software and upgrades.
d) Secure password and file protection capabilities