Differential Scanning Calorimeter

Closed Solicitation Posted

This opportunity was awarded. See the award notice from .

Solicitation number
1333ND21QNB640565
Agency
National Institute of Standards and Technology Department of Commerce
Responses due
Set-aside
No set-aside

Opportunity facts

NAICS code
334516 Analytical Laboratory Instrument Manufacturing
PSC
6640 Laboratory Equipment And Supplies

Notice details come from SAM.gov. Updated .

Notice text

2 versions

Update #2 · Latest ·

AMENDMENT A001

The above referenced solicitation is hereby amended to respond to questions submitted.

The due date and time for quotations, September 8, 2021 at 3:30 PM Eastern Time, remains unchanged by reason of this modification.

The following question and response are hereby provided.

QUESTION:  Upon closer review, the specifications are written in a specific way to lock out other manufacturers. Will  they allow or accept innovation differences for a guarantee of better performance? Specifications in question are #7, #8, and #10f.

RESPONSE:   It is the technical opinion of the Government that the specifications are not lock out specifications. NIST has a scientific basis for each specification. NIST will review all quotations received. Quoters must provide a detailed technical explanation of their “guarantee of better performance”. A statement that the quoter guarantees that what they have is better, on its own, shall not be sufficient.

Relevant to specification #7. If the quoter has another design, NIST is willing to review a detailed technical explanation of how this temperature measurement would be performed without a thermocouple. It would need to be in the same vicinity as the samples to ensure that the measurement was being made in the same location.

Relevant to specification #8. The requirement for “user-exchangeable” is for a couple of reasons. One reason would be, performing oxidation induction temperature measurements, which often foul the cell. This could also be for performing experiments such as curing kinetics, where there is a good chance that something will off-gas into the cell and potentially contaminate it.

Relevant to specification #10f. The question about this specification is not clear. NIST notes that when performing modulated DSC, NIST requires the temperature modulation in the output file.                                                                                                                                                                     END OF AMENDMENT A001 TEXT

The Contractor shall provide a Differential Scanning Calorimeter (DSC) including installation, training and warranty.  

Reference the attached documents for the complete solicitation. 

Update #1 ·

The Contractor shall provide a Differential Scanning Calorimeter (DSC) including installation, training and warranty.  

Reference the attached documents for the complete solicitation. 

Attachments

Files attached to this notice, newest first
File Type Posted
SYN SOL.docx DOCX document
FINAL SOW FOR RFQ.pdf PDF

Notice history

Notices posted for this opportunity, newest first
Notice Type Posted
Differential Scanning Calorimeter Award Award Notice
Differential Scanning Calorimeter This notice · Latest solicitation Solicitation

On GovTribe

Work this opportunity on GovTribe

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity