Blood Irradiator

Awarded Award Notice Posted

Solicitation number
HT0014-16-R-0001
Agency
Defense Health Agency Department of Defense
Awarded
to Rad Source Technologies, Inc.
Set-aside
No set-aside

Opportunity facts

Contract number
HT001416C0006 Federal contract award
NAICS code
334517 Irradiation Apparatus Manufacturing
PSC
Not on record
Place of performance
Walter Reed National Military Medical Center (Wrnmmc) 8901 Wisconsin Avenue Bethesda, Maryland 20889, United States

Notice details come from SAM.gov. Updated .

Notice text

Added: Dec 31, 2015 1:08 pm The vendor shall provide a blood product irradiator with the following minimum physical, functional and performance characteristics: (a) The irradiator must be
designed and FDA 510(k) approved as a medical device to irradiate blood components to prevent Transfusion Associated - Graft Versus Host Disease (TAGVHD), (b) The irradiator must deliver a dose of irradiation of 2500 cGy (25 Gy) targeted to the central portion of the blood product container and 1500 cGy (15 Gy) should be the minimum dose at any other point. This must be confirmed with dosimetry testing at the time of installation and six (6) months after installation, (c) The irradiation source of the irradiator should be by X-ray tubes and not contain radioactive source material (e.g., cesium-137 or cobalt-60), (d) The irradiator must
weigh no more than 2,200 pounds (1,000Kg) in order to comply with the facility's weight-bearing load limitations.

The vendor shall provide annual servicing of this equipment for a period of five (5) years.

Attachments

Files attached to this notice, newest first
File Type Posted
HT0014-16-R-0001.pdf PDF

Notice history

Notices posted for this opportunity, newest first
Notice Type Posted
Blood Irradiator This notice · Latest Award Notice
Blood Irradiator Original Award Notice

On GovTribe

Work this opportunity on GovTribe

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity