Specifications Project Title: Walk-In Biological Safety Cabinets
DRD #: 2750
Specifications required for this procurement include:
1. Walk-In style Class II A2 Biological Containment Enclosure
2. Equal in quality to Brand Name: NuAire AutoLabgard or Baker Company BioPROtect III.
3. Single Door Access
4. Interior working volume (nominal) no less than 160cu.ft.
5. Interior height 74 inches or greater
6. Interior depth 34 inches or greater
7. Supply and exhaust HEPA filters rated at 99.99% efficiency
8. Exhaust into the room
9. Inside constructed of stainless steel/outside constructed of stainless or cold-rolled steel
10. Power 220V, 20A, 60Hz to power blowers.
11. Two 115V, 50/60Hz connections to duplex electrical outlets [115V] which shall be located on the interior side walls (one each side wall) of the cabinet.
12. Fluorescent lighting to provide illumination to the interior
13. Controls for the motors and fluorescent lights should be located on the front of the enclosure along with differential pressure gauge
14. Exhausts into laboratory
15. Two duplex electrical outlets [115V] shall be located on the interior side walls (one each side wall).
16. Dual doors each with an air capture grille
17. Stainless steel cart with a work surface with 21sq ft. minimum.