Optical Spectrum Analyzer (OSA) Technical Specifications

21 June 2018

These specifications represent the minimum requirements for the device.

1. Display: The device must have a display which may be integrated into the device but may take the form of a laptop with associated software. If a laptop is used, it MUST be included with the OSA purchase.

2. Fiber Optic Input: FC/PC or FC/UPC is preferred.

3. Free Space Input: The device must be capable of free space optical input.

4. Wavelength: 600 – 1700 nm. Must be able to measure 850, 1300, 1310, and 1550 nm sources.

5. Level Sensitivity: < - 50 dBm.

6. Spectral Resolution: 7.5 GHz.

7. Spectral Accuracy: ±2 ppm.

8. Spectral Precision: 1 ppm.

9. Input power (CW): > 3 dBm