STATEMENT OF WORK.pdf
PDF 117 KB Posted
- Attached to
- OL 750D SERIES SCANNING SPECTRORADIOMETER Federal contract opportunity
- Solicitation number
- 80NSSC220659
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work and Requirements for an Automated Spectroradiometer
Workstation for Source Spectral Irradiance Measurements
General requirements
Automated double monochromator‐based spectroradiometric system for characterizing laboratory sources and instruments in the near‐UV to near‐Infrared spectral regions
Must have sources and accessories for performing NIST‐traceable system response calibrations
Integrated application software for computer‐controlled data acquisition
Modular system capable of upgrades and interchanging accessories
Must be capable of operating in a FED‐STD‐209E Class 10,000 cleanroom
Optical requirements
Spectral range: 200‐1800nm (0.2 – 1.8μm)
Spectral resolution: 1‐2nm
Gratings blazed for 250 nm (1200 g/mm), 500 nm (1200g/mm), and 1600nm (600 g/mm)
Stray light < 10‐8
Detectors optimized for sensitivity and performance from 200‐1800nm (0.2 – 1.8μm)
PMT detector optimized for the UV
Silicon detector optimized for UV‐VIS
Germanium detector optimized for VIS‐NIR
Integrated order sorting filters
Interchangeable slits
PTFE coated integrating sphere mounted to the entrance port
Technical Contact: Scott Janz, 301‐614‐5987
Date: 4/21/22