This statement of work outlines requirements for a single sample benchtop viscometer to be provided under a Department of the Navy solicitation. The instrument must measure viscosity, density and temperature from a single sample over a range of 0.2 to 200 cSt and 20 to 100 degrees Celsius, with 0.01 degree Celsius temperature accuracy. Delivery and installation are required at the Southwest Regional Maintenance Center in San Diego by 3 November 2023. One year of factory warranty must be included. Onsite training and minimum ten-year replacement part availability are also requirements.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| RFC - N5523623Q0199.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK (revised)
DATE: 7 September 2023
1. SCOPE: SWRMC Code 200 Single Sample Benchtop Viscometer
0. Points of Contact:
SWRMC Code 1150: Ted B. Cook, ted.b.cook.civ@us.navy.mil (619) 556-4362 Code 205B: Brian Harman, brian.c.harmon.civ@us.navy.mil (619-556-0383
1. DESCRIPTION: Vendor shall provide a single sample benchtop Viscometer. That covers the entire viscosity, density and temperature range from a single sample. This instrument will be used for measuring viscosity of transparent and dark/opaque lubricating oils, hydraulic oils, and other petrochemical fluids per ASTM D445. Similar to the SVM 3001 Viscometer Manufacturer: Anton Paar with the following specifications:
1. The viscometer shall have the dimensions that allow it to fit on a standard laboratory bench. Foot print of no more than 30” x 36”.
1. The viscometer must provide a single-point temperature calibrations, and must include a high precision digital thermometer and probe.
1. The viscometer shall provide fully automated self-cleaning.
1.3. Minimum viscosity range 0.2 cSt to 200 cSt.
1.4. Temperature range 20C to 100C +/- 0.01C
1.5. Viscometer must be provided with factory calibration in an ISO 9001-registered, A2LA Accredited laboratory.
1.6. Replacement parts must be available for a minimum of 10 years after date of sale.
1.7. Viscometer shall only have external connections for power of 110VAC, 50-60Hz. No additional facility connections shall be required.
1.8. On-site training shall be included along with delivery and installation of the instrument.
2.10 A factory warranty for a minimum of 1 year must be included at no additional charge.
2. DELIVERY LOCATION:
2.1. Southwest Regional Maintenance Center (SWRMC)
3755 Brinser Street BLDG 61 San Diego, CA 92136 Point of Contact: Ted B. Cook, (619) 556-4362, ted.b.cook@navy.mil
3. PERIOD OF PERFORMANCE DATES:
3.1. Upon Award to 3 November 2023.