Specifications for Gel Imaging System.docx
DOCX document 16 KB Posted
- Attached to
- Gel & Blot Imaging System Federal contract opportunity
- Solicitation number
- 12405B24Q0369
About this file
This document is a Request for Quote (RFQ) for the procurement of a Gel & Blot Imaging System. The RFQ specifies the technical requirements for the system, including hardware/physical requirements, image acquisition requirements, and imaging application requirements. The government is seeking a benchtop integrated gel and blot imaging system with features such as a touchscreen interface, automatic sample rotation, LED lights, high resolution CCD camera, and multiplex fluorescence capability.
The RFQ is issued by the Department of Agriculture Agricultural Research Service, Southeast Area, with a solicitation number of 12405B24Q0369. The contract is anticipated to be a Firm Fixed Price award, with evaluation criteria focusing on technical acceptance, past performance, and price. Quotes are due by August 8, 2024, and the government will consider all timely submissions using a best value tradeoff approach. The associated NAICS code is 334516 (Analytical Laboratory Instrument Manufacturing) with a small business size standard of 1,000 employees.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Commercial Items Provisions 1134603.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications for Gel and Blot Imaging System
Hardware/physical requirements:
· Integrated gel and blot imaging system
· Touchscreen interface
· USB port
· Imaging area at least 17 cm x 22 cm
· Benchtop footprint less than 40 cm x 70 cm
· Automatic sample rotation (mechanical rotation of sample stage)
· LED lights in transilluminators (instead of UV lights)
· Storage requirements: >200 GB Image acquisition requirements:
· High resolution CCD camera, > 9.0 megapixel
· Autoexposure
· Autofocus
· Autozoom, up to 8x zoom Imaging application requirements:
· Protein gels (colorimetric and fluorescence signals)
· Nucleic acid gels (ethidium bromide and fluorescent stain signals)
· Colorimetric western blots
· Chemiluminescence western blots (horseradish peroxidase and alkaline phosphatase substrates)
· Fluorescent western blots (fluorescence from visible and near-infrared fluorophores)
· Multiplex fluorescence capability
· Custom mode to image with multiple signals
File details come from the government source that posted it. Updated .