This document is a Request for Quote (RFQ) for the procurement of a Benchtop MRI System. The key specifications for the benchtop MRI system include a 0.55 Tesla permanent magnet, gradient strength over 250 mT/m, ability to measure T2 relaxation data and collect MRI images of meat samples, and software for data acquisition and analysis. The solicitation is issued by the USDA Agricultural Research Service, Southeast Area, with a NAICS code of 334516 (Analytical Laboratory Instrument Manufacturing). The Government anticipates awarding a Firm Fixed Price contract to the responsible offeror whose quote conforms to the RFQ and specifications, using a tradeoff between technical acceptance, past performance, and price. Quotes must be received by July 19, 2024, 4:00 pm Central Standard Time. Questions are due by July 17, 2024, 3:00 pm Central Standard Time.

View the file

Other files for this federal contract opportunity

Other files attached to Benchtop MRI System, newest first.
File Type Posted
Commercial Items Provisions 1134602.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

A benchtop MRI is needed with following specifications:

· Benchtop configuration

· Able to measure T2 relaxation data of meat samples

· Able to perform T2 relaxation distribution analysis of meat samples

· Able to collect MRI images of meat samples

· 0.55 Tesla permanent magnet

· Gradient strength (built-in gradient amplifier): > 250 mT/m (x,y); 350 mT/m (z-direction)

· Gradient strength (external DC-600 gradient amplifier): ≥ 1200 mT/m (x,y); 1500 mT/m (z-direction)

· Gradient efficiency: > 250 mT/m/A (x,y); 330 mT/m/A (z-direction)

· 20 mm probe exchangeable with smaller diameters

· Sample diameter imaging20 mm
· Field homogeneity (10 mm)< 10 ppm
· Field homogeneity (5 mm)< 3 ppm
· Spectral resolution (5 mm)< 1 ppm

· Temperature programing capacity (from ambient temperature to 80C at least)

· Software to acquire data of NMR and MRI and Matlab software to process and analyze data

· Open MATLAB interface

File details come from the government source that posted it. Updated .