This solicitation seeks scientific test equipment including vector network analyzers, frequency offset measurement licenses, calibration modules, handheld microwave analyzers, signal generators, and signal analyzers. Key requirements include vector network analyzers with 4 ports and a frequency range of 10 MHz to 50 GHz, frequency offset measurement licenses, electronic calibration modules with a range of 10 kHz to 18 GHz, handheld microwave analyzers with built-in GPS and a range of 10 kHz to 18 GHz, signal generators with a range of 10 kHz to 3 GHz and less than 10^-7/year frequency aging, and a signal analyzer with a range of 10 Hz to 3 GHz and less than 10^-7/year frequency aging. The solicitation was issued by the Department of the Navy Naval Sea Systems Command to acquire this scientific test equipment.

View the file

Other files for this federal contract opportunity

Other files attached to Scientific Test Equipment, newest first.
File Type Posted
Combined Synopsis Solicitation.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

CLIN MINIMUM SPECIFICATIONS/SALIENT FEATURES QUANTITY

0001 Vector Network Analyzer

• 4-port

• 50 ohm

• Direct access to source/receiver ports

• Frequency range: 10 MHz to 50 GHz

• Dynamic range: at least 130 dB typical

• Noise floor: less than -120 dBm typical

• Sweep time: less than 0.01 ms/point

• Accessibility: display, USB, ethernet, SCPI commands

• Frequency offset measurement capability

• IF input and mixer switch capability

1 EACH

0002 Frequency-offset measurements Node-locked Perpetual License for Vector Network Analyzer

• 12 month software support subscription

1 EACH

0003 Calibration Module

• Electronic calibration module supported by current Keysight network analyzers

• 2-port

• 50 ohm, N-female

• Frequency range: 10 kHz to 18 GHz

2 EACH

0004 Handheld Microwave Analyzer

• 2-port

• 50 ohm, N-female

• Vector network analyzer and spectrum analyzer functionality

• Supports current Keysight electronic calibration modules

• Frequency range: 10 kHz to 18 GHz

• Output power: -40 to 0 dBm typical

2 EACH

• Output power accuracy: less than 0.5 dB typical

• Calibrated transmission and reflection uncertainty: less than 0.2 dB / 2 degrees typical

• Spectrum analyzer input accuracy: less than 0.3 dB typical

• Noise floor: less than -160 dBm typical

• Dynamic range: at least 100 dB typical

• Independent source

• Zero span sweep time: 1 microsecond to 1 hour

• Real-time spectrum analysis bandwidth: at least 10 MHz

• GPS receiver built-in for frequency reference

• DC bias-tee output capability of at least 30 V and 250 mA

• Accessibility: display, USB, ethernet, SCPI commands

• Handheld with battery (at least 2 hours)

• Operating temperature: -10 to +50 degrees Celsius

0005 Signal Generator

• 50 ohm, N-female

• Frequency range: 10 kHz to 3 GHz

• Frequency aging: less than 10^-7/year

• Output power: -120 to +10 dBm

• Output power accuracy: less than 1 dB typical

• Accessibility: display, USB, ethernet, SCPI commands

3 EACH

0006 Signal Analyzer

• 50 ohm, N-female

• Frequency range: 10 Hz (not kHz) to 3 GHz

• Frequency aging: less than 10^-7/year

• Input attenuation: 0 to 80 dB in 1 dB steps

• Amplitude accuracy: less than 0.3 dB typical

• Noise floor: less than -170 dBm typical

1 EACH

• Zero span sweep time: 1 microsecond to 1 hour

• Real-time spectrum analysis bandwidth: at least 10 MHz

• Accessibility: display, USB, ethernet, SCPI commands

File details come from the government source that posted it. Updated .