SPECIFICATIONS FOR Micro Distillation Analyzer The following is a list of requirements for purchase of a micro distillation analyzer designed primarily for the determination of quantitative boiling range of petroleum fluids (e.g. fuels). This instrument will be used to test for boiling range (ASTM D86-19) of petroleum samples at the Air Force Petroleum Agency’s Aerospace Fuels Laboratories at Wright Patterson AFB OH.

1. This instrument must be an automated distillation instrument that is fully compliant with ASTM D86-19, “Standard Test Method for Distillation of Petroleum Products and Liquid Fuels at Atmospheric Pressure”.

2. Must be capable of providing analysis of at least 5 samples per hour with minimal operator time per test (5-10 min).

3. Must be have a level follower/recording mechanism that has a resolution of 0.1% volume or better with a maximum error of 0.3% volume between the 5% and 100% range.

4. Be capable of providing results in appropriate units (mL & oC) using analyzer software, that meets the precision and bias requirements specified in the method.

5. Be capable of operating at temperature ranges of 0o to 400oC with a sensitivity of ± 0.1oC.

6. Provide integrated user controls for operation, calibration, and standardization.

7. Include a printer to record/print results.

8. Must be supplied with all accessories required for installation and operation.

9. Must operate on standard 110-120V, 60Hz

10. Size shall not exceed 20in x 20in x 20 in. (l x w x h)

11. Must be delivered on site within 60 days of contract

a. Must be delivered with instrument manuals for setup and operation

Note: Copies of ASTM D86-19 may be obtained by writing; ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959; by telephone at 610-832-9255; by e-mail at service@astm.org; or through the ASTM website at www.astm.org