SPECIFICATIONS.docx

DOCX document 17 KB Posted

Attached to
Multichannel Spectrophotometer for Corvallis, OR Federal contract opportunity
Solicitation number
12905B24Q0034
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area

About this file

This solicitation requests quotes for a multichannel spectrophotometer. The required product must be able to quantitate at least eight samples in less than 20 seconds or a 96-well plate in under six minutes, operate within a 190-850nm wavelength range, integrate with LIMS software, identify contaminants using built-in algorithms, detect up to 10,000ng/dL dsDNA or 145mg/mL IgG, use small sample volumes of 1-2uL, and have a wavelength accuracy of +/-1nm. Quotes are due by February 10, 2024 at 12:00pm PST and should be submitted to Brian.Fien@usda.gov. The award will be made to the responsible offeror submitting a technically acceptable quote at the lowest evaluated price. The product must be delivered within 60 days of award to 3420 NW Orchard Ave, Corvallis, OR 97330. The soliciting agency is the US Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area.

View the file

Other files for this federal contract opportunity

Other files attached to Multichannel Spectrophotometer for Corvallis, OR, newest first.
File Type Posted
Terms and Conditions (2).docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Multichannel Spectrophotometer for 3420 NW Orchard Ave, Corvallis, OR 97330

Minimum Specifications

- High throughput ability: Quantitate at least eight samples in less than 20 seconds or a 96-well plate (eight samples at a time) in less than 6 minutes.

- Wave Length: 190–850 nm

- LIMS ready

- Contaminant Identification: Built-in Acclaro Software algorithms

- Detection limits: Up to 10,000 ng/µL dsDNA or 145 mg/mL IgG

- Smaller sample volumes: 1 - 2 uL,

- Wavelength acccuracy: +/- 1 nm

- Ambidextrous pipetting – the pipette guide should be movable from side-to-side to allow both left- and right-handed user