Requirements / Specification Document
TITLE: WR-6 Extender Heads
Requesting Laboratory / Division / Group:
Communications Technology Laboratory Radio Frequency Technology Division Guided Wave Electromagnetics Group (672.01)
General Statement of Need:
The National Institute of Standards and Technology (NIST)/Communications Technology Laboratory (CTL) enables science and industry by providing essential measurement methods, instrumentation, and standards to support all phases of communications technology development. As part of this effort, the RF Technology Division of the Communications Technology Laboratory develops and maintains traceability paths for scattering parameter and power calibrations. In order to pursue these activities, the RF technology division requires vector-network-analyzer frequency-extender heads and calibration kits for various rectangular waveguide bands.
Items and Ancillary Services:
Item 1: WR-6.5 Extender Head Quantity: 2 Specifications:
1.1 Shall extend the frequency range of S-parameter and wave-parameter measurements for PNA-X and ZVA vector network analyzers.
1.2 Shall have an operating frequency range that spans at least 110 GHz to 170 GHz.
1.3 Shall have a WR-6.5 test port flange. Test port flange shall be a precision flange that includes both an anti-cocking feature and bores for slip-fit fine-alignment dowels.
1.4 Shall have both transmit and receive capabilities.
1.5 Shall have an integrated manually adjustable attenuator with a minimum attenuation of 0 dB and a maximum attenuation of at least 25 dB.
1.6 Shall have an output power of at least +9 dBm across the entire band from 110 GHz to 170 GHz.
1.7 Shall have a minimum dynamic range of at least 100 dB across the entire band from 110 GHz to 170 GHz.
1.8 Shall have a typical power magnitude stability equal to or better than ± 0.3 dB (10 Hz IFBW after 1 h warm up).
1.9 Shall have a typical phase stability equal to or better than ± 4 degrees (10 Hz IFBW after 1 h warm up).
1.10 Shall have a typical raw coupler directivity of at least 25 dB across the entire band from 110 GHz to 170 GHz.
1.11 Shall have a typical test port input power damage limit of at least 20 dBm.
1.12 Shall include one (1) cable set for each extender head. Cable set shall include at least (1) RF, (1) LO and (2) IF cables for each extender head. The length of the cables shall be 1.2 m or 48 inches.
Item 2: WR-6.5 Calibration Kit Quantity: 1 Specifications: 2.1 All calibration kit standards shall have WR-6.5 test port flanges and operate in the frequency range from 110 GHz to 170 GHz.
2.2 Shall include at least one flush short circuit standard.
2.3 Shall include at least one fixed matched load standard with precision flange that includes both an anti-cocking feature and bores for slip-fit fine-alignment dowels.
2.4 Shall include at least one quarter-wave shim with bores for slip-fit fine-alignment dowels.
Service Requirements / Item Specifications:
Section Requirements/Specifications Planning Considerations ☒ N/A
☐ List any visits required to allow potential vendors to see site specific constraints.
[Explain Requirements.] ☐ US citizen required on-site (foreign nationals require pre-registration 30 days prior to visit.)
☐ Vendor Representative On-Site more than three (3) days.
☐ Other Considerations:
[List other considerations to ensure an accurate quote.] Shipping & Delivery ☐ N/A
Shipping Address:
☒ NIST Boulder
ATTN: Christian Long MAIL Stop 672.01 325 Broadway Boulder, CO 80305
☐ [Alternate Shipping Address] Shipping Criteria:
☒ Partial Delivery Acceptable
☐ [Enter Shipping and Delivery information by item.] ☐ Direct Delivery to Building Required
[Explain Need for Requirement] ☐ Building has Loading Dock
Containerization Preference: [Enter Container Type] Other Requirements:
[List Any Other Requirement] Delivery Date Criteria:
☒ Delivery of Goods, Installation, and Training shall be competed no later than: 24 weeks After Receipt of Order (ARO) ☐ Delivery No Earlier Than: Click or tap to enter a date.
☐ Other:
[List Other Shipping, Delivery or Special Requirements.] Electronic Media ☒ N/A
[Enter Electronic Media Requirements by Item.]
☐ Software ☐ Electronic Manuals ☐ License Key ☐ Other:
[List Any Other Electronic Media Requirements] ☐ Email address for Software Delivery:
[Enter Address] Installation ☒ N/A
[Enter Installation Requirements by Item]
☐ Vendor Technician on Site ☐ Rigging ☐ Uncrating / Unpackaging ☐ Removal of Packaging Material ☐ Equipment Set Up ☐ Start-Up Services ☐ Turnkey Installation ☐ Other:
[Enter Details]
NOTE:
When installation is required, TPOC shall be the coordination point.
Facility / Utility Considerations ☒ N/A
[List Site Specific Utility Considerations by Item] ☐ OFPM Work Order Numbers
[List Work Order Numbers] ☐ Power Available Voltage: [Enter voltage] (V) Maximum Current: [Enter max current draw] (A) Phase: [Choose Phase] Frequency: [Typically 60Hz] (Hz) NEMA Plug Type: [Enter Plug Type] (i.e.: NEMA L5-20P) Environment: [Choose Environment] Other Power Considerations:
[List Any Other Power Considerations.] ☐ Utility Connections:
[List Relevant Available Utilities, Pressure(s) and Flowrate(s).] ☐ Other Considerations:
[List Other Considerations to Ensure an Accurate Quote.] Warranty ☐ N/A
Base Manufacturer Warranty is acceptable
[If you require an Extended Manufacturer Warranty, describe the warranty requirements per item.]
What is included?
☒ Labor ☒ Parts ☒ Travel ☒ Cost and Liability for Returns ☐ Other [Enter Other Requirements] ☒ Warranty Length: 1- year
Training ☒ N/A
[Training Requirements (Who, When, What, Where)] ☐ On-Site
Is Travel Required: [Choose Answer]
☐ Off-Site Location: [Enter Location(s)] ☐ Number of Trainees: [Enter Number of Trainees] Scope of Training ☐ Operations ☐ Maintenance ☐ Troubleshooting ☐ Safety ☐ Other [Enter Other Scope Requirements] ☐ Training materials to be provided:
[Describe Requirements] ☐ Other [Enter Other Training Requirements] NOTE: When training is required, TPOC shall be coordinator.
Inspection and Acceptance ☐ N/A
[Inspection Requirements Per Item (Who, When, What, Where)] ☒ On-Site
[Inspection Requirements Per Item] ☐ Government Acceptance Expected to Not to Exceed 7 Business Days ☒ Government Acceptance Expected to Exceed 7 Business Days 14 Calendar Days ☐ Inspection and Acceptance Plan:
[Explanation] ☐ Other:
[Enter Other Requirements] Maintenance ☒ N/A
[Describe Maintenance Requirements Per Item] Service Plan Length: [i.e. 3 - Years]
What is Included?
☐ Labor ☐ Spare Parts ☐ Consumables ☐ Travel
☐ Cost and Liability for Returns ☐ Software Updates ☐ Manuals l
☐ Other [Enter Other Requirements] ☐ Requirements:
[Enter Requirements] (i.e.: On-Site Within 72 hours, etc.)
☐ 1 Option Period ☐ 2 Option Periods ☐ 3 Option Periods ☐ 4 Option Periods (Maximum) Government Furnished Material / Equipment ☒ N/A
[List Any Government Furnished Material / Equipment Per Item] ☐ Samples for Testing
☐ Government Equipment which will leave Government Site (TPOC must coordinate with Division Property Custodian)
Description: [Enter Description] Serial Number: [Enter Serial Number]
NIST Property Number: [Enter NIST Property Number] ☐ Other:
[Enter Other Details] Travel ☒ N/A
[Describe any Travel Requirements Per Item] ☐ Explain:
[Enter Justification / Requirement] ☐ Other:
[Enter Other Requirements] NOTE TO VENDOR: To the maximum extent possible, include travel requirement estimates as part of pertinent sections above (Installation, Maintenance, Repair, etc.). Otherwise, Travel will be a separate line item, reimbursable per the Federal Travel Regulation.