Salient Characteristics
a) Environmental DNA pump systems (14 with backpack frames, 2 without)
b) Should include manual and automatic settings for flow rate (L/min), Maximum pressure settings
(PIS), sample volume (L) and with an alarm to alert users of filter clogging.
c) Units must include onboard GPS for global positioning with data recording during sample collection. Data should be retrievable by USB or Bluetooth.
d) Pumps must connect to silicone tubing with an integrated sampling pole capable of collecting replicate samples, up to 3, simultaneously (14 of the 16 units). Two of the units must have telescoping bipod or tripod stabilized sampling poles to facilitate mid‐stream sample collection from small to mid‐sized rivers.
e) Units must include (for 14 of the 16 units) an integrated backpack frame to facilitate backcountry applications.
f) Units must be powered by lightweight, lithium batteries.
g) Units must include USB ports for data retrieval, software updates.