REPLACEMENT STEAM STERILIZER SYSTEM

STATEMENT OF WORK (SOW)

1.0 BACKGROUND

The Naval Surface Warfare Center, Dahlgren Division (NSWCDD), has a requirement for a replacement steam sterilizer system to be used in the NSWCDD . A steam sterilizer (also called autoclave) is an essential component to the safety and cleanliness of our bio-safety facility and associated operations. Currently, the existing steam sterilizer system in lab that requires replacement has consistently gone down for repairs, has proven unreliable, and has led to many work stoppages and schedule delays.

2.0 REQUIREMENTS

Original Equipment Manufacturer (OEM) Specifications

3.0 REQUIREMENTS

3.1 The Contractor shall provide one (1) each Tuttnauer™ Model 66C-

SPD1-V-CS Pre-Vac Steam Sterilizer, part number 66C-SPD1-V-CS, manufactured by Tuttnauer U.S.A. CO., LTD., that meets the following specifications:

3.1.1 Steam Sterilizer unit must be Stainless Steel, preferably

SS316, but minimum SS304 to work with Reverse Osmosis water.

Stainless Steel required due to the Reverse Osmosis “clean” water that is required for the Tissue Culture capability.

3.1.2 Steam Sterilizer unit steam process plumbing, including Steam

Generator, pipes, valves, and fittings must be SS316.

3.1.3 Steam Sterilizer unit must be double door pass-through with

Bio-Seal. Doors must be vertical sliding doors due to facility space constraints.

3.1.4 Steam Sterilizer must have internal dimensions of 26W x 26H x

39D and have a capacity of 316L.

3.1.5 Steam Sterilizer unit controls: Primary controls must be located on the BSL-2 end on the right-hand side of the unit.

Secondary controls must be located on the BSL-3 end on the left-hand side of the unit.

3.1.6 Steam Sterilizer unit must have gravity, liquid, pre-vac, effluent decontamination cycles. Test Cycles: Vacuum Leak Test, Bowie Dick Test, Effluent Sterilization Bio-Filter Cycle (for BSL3 applications)

3.1.7 Steam Sterilizer unit must have impact printer for cycle information readouts on BSL-2 side.

3.1.8 Steam Sterilizer unit must have all components

accessible/serviceable on right-hand side of BSL-2 end.

3.1.9 Steam Sterilizer unit must have Bio-Seal consisting of

Stainless Steel Bio-Seal Frame and Gasket.

3.2 The Contractor shall delivery the unit to the NSWCDD located in

Dahlgren, Virginia.

3.3 The Contractor shall install the unit in accordance with (IAW)

OEM specifications.

3.4 The Contractor shall test the unit to ensure it is operating

IAW OEM Specifications.

3.5 The Contractor shall coordinate with the technical point of contact (TPOC) identified below to complete this work.

3.6 The Contractor shall provide each Contractor employee with the personal protective equipment required to do their job safely and in compliance with all applicable regulations.

3.6.1 All Contractor personnel shall maintain social distancing and wear a facemask at all times while onsite at the NSWCDD base until notified otherwise by the government.