This statement of work describes requirements for a high/low temperature tribometer procurement by the National Aeronautics and Space Administration Glenn Research Center. The tribometer must be capable of measuring friction and wear for reciprocating, oscillating, and rotational motion between wear couples. Key requirements include rotational speeds up to 2000 rpm, a maximum wear track radius of 40 mm, normal loads up to 60 N, friction forces up to 20 N, rotating temperatures from 0 to 150 C, linear motion up to 60 mm, linear motion temperatures from -100 to 450 C, and data acquisition rates up to 400 Hz. The statement of work also requires all ancillary equipment for tribometer operation and precision friction and wear measurement, including active thermal control systems, sample and specimen holders, in-situ calibration equipment, on-site setup, installation, calibration, and two years of on-site preventative maintenance, calibration, adjustments, cleaning, and machine adjustments. A certificate of calibration and accuracy is also required.
The related federal contract opportunity is solicitation number 80NSSC23845323Q for a high/low temperature tribometer from the National Aeronautics and Space Administration Shared Services Center.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work Tribometer
NASA GRC requires the procurement of a tribometer capable of measuring fric�on and wear for reciproca�ng, oscilla�ng, and rota�onal mo�on between the wear couple.
Requirements Shall be:
Rota�onal speeds up to 2000 rpm Maximum wear track radius of 40 mm Maximum normal load of 60 N Maximum fric�on force of 20 N Rota�ng temperature range of 0 to +150 C Linear mo�on range of 60mm Linear mo�on temperature range of -100 C to +450 C Real Time measurement and display of wear and fric�on Data acquisi�on rate up to 400 Hz Requisite materials to maintain calibra�on of instrumenta�on All ancillary equipment required to operate the tribometer and conduct precision fric�on and wear measurement, including but not necessarily limited to: ac�ve thermal control system, sample holders, specimen holders, in situ calibra�on equipment, on-site tribometer setup-installa�on-calibra�on, 2 years of on-site preven�ve maintenance-calibra�on-adjustments-cleaning-machine adjustments- etc… Cer�ficate of calibra�on and accuracy