1. Objective/Requirements

Code 564 has several oscilloscopes that are antiquated, and in some cases presenting with limited functionality due to their age. Many of the oscilloscopes don’t have USB ports to save and retrieve testing data.

The requirements for this procurement are:

Acquire an oscilloscope that has at least 2.5GHz of bandwidth capability Has USB ports available to collect and store test data Has GPIB adapter for the use of automating the scope for specific test plans Has at least four channels Able to support the use of differential, active, and current probes.

i. Differential probes have threshold of at least +/- 8V

ii. Current probes can handle at least 30A of peak current

Is software upgradable to include analysis kits

i. Jitter analysis kit

2. Characteristics, Scope, and Specs

Waverunner 9254 oscilloscope:

o Baseline Specs:

2.5GHz Front-End capability 20GS/Sec 4 input channels o Software Upgrades

WR9K-JITKIT

Clock, Clock-data Jitter Analysis and Views of time, statistical, spectral, and jitter overlay o External connections

USB 2.0

GPIB IEEE 488.2 adapter o Probes ZS2500: 2.5GHZ 0.9pF 1MOhm Active Probe ZD1500: 1.5GHz 1.0pF Active Differential Probe, +/- 8V CP031: 30A 100MHz Current Probe – AC/DC, 30A rms, 50A peak pulse, 1.5-meter cable

3. Place of Performance

This oscilloscope will be used in Code 564, building 11.

4. Period of Performance This scope is expected to perform over the next 5 years.