SATPC0039622 Tab 04 4 Specification.pdf

PDF 22 KB Posted

Attached to
Cryocooler for VF-15 AL230 Cryocooler Federal contract opportunity
Solicitation number
80NSSC25913703Q
Issued by
National Aeronautics and Space Administration Shared Services Center

About this file

This document is a Parameter Specification for procuring a cryocooler for the VF-15 vessel at NASA Glenn Research Center (GRC), dated 7-24-2025. The specification outlines detailed technical requirements for the cryocooler, including a base temperature below 20 K, heat lift capabilities of over 20 W at 20 K and over 40 W at 30 K, and a vertical orientation with the cold tip positioned on top. Additional requirements specify an air-cooled compressor system, mounting via a Conflat flange (6" or 8" OD preferred), and a power supply of 208 V, 3 Phase, 60 Hz.

View the file

Other files for this federal contract opportunity

Other files attached to Cryocooler for VF-15 AL230 Cryocooler, newest first.
File Type Posted
SATPC0039622 Tab 07 Capability Statement Commerical.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for the procurement of a cryocooler for the VF-15 vessel at NASA GRC

7-24-2025

Parameter Specification

Base temperature < 20 K

Heat lift > 20 W at 20 K, > 40 W at 30 K

Cryocooler orientation vertical with cold tip on top (cryocooler pointed up)

Compressor cooling system air cooled

Mount type and size Conflat flange, preferably 6” or 8” OD

Allowable supply voltage 208 V, 3 Phase, 60 Hz