This Statement of Work (SOW) outlines the requirements for a portable helium leak detector to be delivered to Marshall Space Flight Center in Huntsville, AL. The required equipment must be a mass spectrometer-based detector with an integrated dry pump system capable of pumping speeds of at least 3 m³/hr during pumpdown and 2.5 l/s helium pumping speed in fine mode.

The technical specifications include dual mode operation (vacuum and sniffer), sensitivity requirements of 5E-12 mbarl/s in vacuum mode and 7E-9 mbarl/s in sniffing mode or better, KF/NF 25 connection port, integrated user control interface, and 115V/60 Hz power operation. The period of performance is scheduled for January 2025 through April 2025. This procurement is being managed by NASA's Shared Services Center for delivery to Building 4605 at Marshall Space Flight Center.

View the file

Other files for this federal contract opportunity

Other files attached to Helium leak detector, newest first.
File Type Posted
SATPC0037409 Tab 10 Question and answer.pdf PDF
SATPC0037409 Tab 10 Open Market RFQ.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SOW ‐ Helium Leak Detector Requirements

Required Characteristics:

• Portable helium leak detector o Mass spectrometer based He detection

• Integrated pump (does not require external backing pump) o Dry pump required (not “wet” or oil‐filled) o Pumping speed of at least 3 m^3/hr during pumpdown o He pumping speed of at least 2.5 l/s in fine mode

• Dual mode (vacuum opera�on and sniffer mode)

• Sensitivity o 5E‐12 mbar*l/s in vacuum opera�on mode (or better) o 7E‐9 mbar*l/s in sniffing mode (or better)

• KF/NF 25 connection port

• Integrated user control interface

• 115V, 60 Hz operation

Shipping Address:

Marshall Space Flight Center

Bldg 4605, Room 8

ATTN: Erin Hayward

Huntsville, AL 35812

Period of Performance: 1/2025 – 4/2025