This document is a Statement of Work (SOW) for a data acquisition system for high-speed Electrical Impedance Tomography (EIT) to be purchased by NASA Ames Research Center. The key requirements include:

The data system must have at least 32 input channels, 16-bit or higher digitization resolution, over 2 MHz bandwidth, and be programmable for complex EIT acquisition sequences. It must include a graphical user interface (GUI) software and a Python API to control data acquisition parameters and stream data. The system will be initially installed at the NASA Ames PLASMA Lab but may be moved to other testing facilities as needed. Delivery is estimated at 5 weeks.

This SOW is associated with a federal contract opportunity (Solicitation Number 80NSSC24877132Q-R1) for a "Spectrum digitizer NETBOX, 16 bit multi-purpose digitizer". The procurement is a brand name requirement from Spectrum Instrumentation GmbH, and is set aside for small businesses. The National Aeronautics and Space Administration (NASA) Shared Services Center is the participating federal agency.

View the file

Other files for this federal contract opportunity

Other files attached to Spectrum digitizer NETBOX, 16 bit multi-purpose digitizer, newest first.
File Type Posted
SATPC0035841 Tab 06 BNJ_Redacted.pdf PDF
SATPC0035841 Tab 10 RFQ Open Market.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

1. Objective/Requirements

NASA Ames Research Center plans to purchase a data acquisition system for high-speed, Electrical Impedance Tomography (EIT) to develop a system capable of time-resolved 3D tomographic reconstruction of the electrical impedance of a sample. One application of this technology is in-situ crack detection in thermal protection materials in high temperature test environments.

The data system will be used to collect simultaneous high frequency (1-10 MHz bandwidth) data from 32 channels suitable for digital demodulation of multiple mixed kHz frequencies.

The data system should be programmable for the complex acquisition sequences required for EIT and with high bit resolution (>=16 bit).

The information listed below is required to complete this purchase.

2. Characteristics, Scope, and Specs

LINE PART NO. DESCRIPTION QTY.

1 DN6.592-32 digitizerNETBOX 16 bit LAN digitizer

Ethernet/LXI interface, BNC connectors, 32 channels 20 MS/s, Single Ended/Differential Inputs,10 MHz bandwidth, 4x512 MSample memory, 230V/115V AC power supply SBench 6 Professional license

2 Cab-Pwr-002 AC power cable for US, Canada and others with NEMA5-15P connector, 180 cm long 1 3 M2p.xxxx-

PulseGen Firmware option: 4 x digital pulse generator adds 4 freely programmable digital pulse generators that use the XIO lines for output

4 DN6.xxx-Rack 19“ rack mounting panel set Rail kit has to be ordered separately at 19 chassis supplier 1 5 DN6.xxx-Emb Extension to Embedded Server:

Intel Quad Core 2 GHz, 4 GByte RAM 128 GByte SSD, full digitizerNETBOX functionality and configuration available, remote Linux command shell (ssh) for user software with access to digitizer, SSD and LAN OpenSuse 12.2 (kernel 4.4.7.)

6 SH1DN6 DDP Shipment for one Netbox DN6 1

The data system shall have the following minimum requirements:

• Simultaneous data acquisition on all channels

• Input channel count >=32

• Digitization resolution >= 16 bit

• Programmable API

• Bandwidth >2 MHz

Instrument Operation and Data Software

• The data system will be supplied with graphical user interface (GUI) software and a Python application programming interface (API) that can perform the following:

• Control data acquisition speed, voltage range, digital outputs, and stream collected data into storage system.

• A GUI software for simple testing

3. Place of Performance

Initial installation of equipment will be at NASA Ames in Code TSM PLASMA Lab but can be moved to other testing facilities as needed.

NASA Ames Research Center, MS234-1 Moffett Field 94085

4. Period of Performance

Delivery of the unit is estimated at 5 weeks not including installation.