SATPC0032016 Tab 04 4 SOW.pdf
PDF 119 KB Posted
- Attached to
- Mixed signal oscilloscope and related hardware Federal contract opportunity
- Solicitation number
- 80NSSC23841596Q
About this file
This statement of work outlines specifications for a mixed signal oscilloscope, differential probe, and current probe to be purchased by NASA Glenn Research Center. The oscilloscope must have 8 analog channels and 64 digital channels, a 1 GHz bandwidth, 6.25 GS/s sample rate, 42V peak input voltage, 100 kΩ input impedance, 2 USB 3.0 ports, rear-panel BNC auxiliary output, and waveform capture rate over 500,000 waveforms/s. It should also include function generator and advanced power measurement capabilities with a 250M/ch maximum record length. The differential probe must have 1500 V differential voltage, 200 MHz bandwidth, 275 V/ns slew rate, and TekVPI interface connectivity. The current probe specifications include a 150 A maximum RMS current, 5 mA minimum sensitivity, 3% DC accuracy, and TekVPI interface connectivity. These products relate to federal contract opportunity number 80NSSC23841596Q issued by the National Aeronautics and Space Administration Shared Services Center to purchase a mixed signal oscilloscope and related hardware.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SATPC0032016 Tab 12 RFQ Open Market.pdf | ||
| SATPC0032016 Tab 12 BNJ_CO_Redacted.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
Background
NASA Glenn Research Center (GRC) is inquiring about purchasing a mixed signal oscilloscope and associated equipment to view waveforms, take measurements, and perform probing on lab equipment.
The oscilloscope must be compatible with the existing TLP058 Probes at GRC and must meet the outlined specifications below.
Oscilloscope Specifications
• Number of Analog Channels: 8
• Maximum number of digital Channels: 64
• Bandwidth: ≥ 1 GHz
• Sample Rate: 6.25 GS/ss
• Absolute Maximum Input Voltage: ± 42V peak
• Input Impedance: 100 kΩ
• USB Interface: 2 USB 3.0 SuperSpeed ports
• Auxiliary Output: Rear-panel BNC Connector
• Waveform capture rate: >500,000 waveforms/s
• Function Generator capabilities
• Advanced Power Measurement and Analysis functionality
• Record Length: 250M/ch maximum
Differential Probe Specifications
• Differential Voltage: ±1500 V
• Bandwidth: ≥200 MHz
• Slew rate: ≥275 V/ns
• Connectivity: TekVPI Interface
Current Probe Specifications
• Max RMS Current: ≥150 A
• Minimum sensitivity: ≤5 mA
• DC Accuracy: ≤3%
• Connectivity: TekVPI Interface
File details come from the government source that posted it. Updated .