SATPC0032016 Tab 04 4 SOW.pdf

PDF 119 KB Posted

Attached to
Mixed signal oscilloscope and related hardware Federal contract opportunity
Solicitation number
80NSSC23841596Q
Issued by
National Aeronautics and Space Administration Shared Services Center

About this file

This statement of work outlines specifications for a mixed signal oscilloscope, differential probe, and current probe to be purchased by NASA Glenn Research Center. The oscilloscope must have 8 analog channels and 64 digital channels, a 1 GHz bandwidth, 6.25 GS/s sample rate, 42V peak input voltage, 100 kΩ input impedance, 2 USB 3.0 ports, rear-panel BNC auxiliary output, and waveform capture rate over 500,000 waveforms/s. It should also include function generator and advanced power measurement capabilities with a 250M/ch maximum record length. The differential probe must have 1500 V differential voltage, 200 MHz bandwidth, 275 V/ns slew rate, and TekVPI interface connectivity. The current probe specifications include a 150 A maximum RMS current, 5 mA minimum sensitivity, 3% DC accuracy, and TekVPI interface connectivity. These products relate to federal contract opportunity number 80NSSC23841596Q issued by the National Aeronautics and Space Administration Shared Services Center to purchase a mixed signal oscilloscope and related hardware.

View the file

Other files for this federal contract opportunity

Other files attached to Mixed signal oscilloscope and related hardware, newest first.
File Type Posted
SATPC0032016 Tab 12 RFQ Open Market.pdf PDF
SATPC0032016 Tab 12 BNJ_CO_Redacted.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

Background

NASA Glenn Research Center (GRC) is inquiring about purchasing a mixed signal oscilloscope and associated equipment to view waveforms, take measurements, and perform probing on lab equipment.

The oscilloscope must be compatible with the existing TLP058 Probes at GRC and must meet the outlined specifications below.

Oscilloscope Specifications

• Number of Analog Channels: 8

• Maximum number of digital Channels: 64

• Bandwidth: ≥ 1 GHz

• Sample Rate: 6.25 GS/ss

• Absolute Maximum Input Voltage: ± 42V peak

• Input Impedance: 100 kΩ

• USB Interface: 2 USB 3.0 SuperSpeed ports

• Auxiliary Output: Rear-panel BNC Connector

• Waveform capture rate: >500,000 waveforms/s

• Function Generator capabilities

• Advanced Power Measurement and Analysis functionality

• Record Length: 250M/ch maximum

Differential Probe Specifications

• Differential Voltage: ±1500 V

• Bandwidth: ≥200 MHz

• Slew rate: ≥275 V/ns

• Connectivity: TekVPI Interface

Current Probe Specifications

• Max RMS Current: ≥150 A

• Minimum sensitivity: ≤5 mA

• DC Accuracy: ≤3%

• Connectivity: TekVPI Interface