SALIENT CHARACTERISTICS.pdf

PDF 440 KB Posted

Attached to
ACQUISITION RECORDER Federal contract opportunity
Solicitation number
SPMYM2-24-Q-2480
Issued by
Defense Logistics Agency Land and Maritime

About this file

This document is a Request for Quote (RFQ) for the procurement of an Acquisition Recorder from A & D Engineering Inc. The solicitation, numbered SPMYM2-24-Q-2480, is being issued by the Defense Logistics Agency Land and Maritime on an unrestricted basis using commercial item procedures. The resultant contract will be a firm fixed price commercial purchase order, with the requirement utilizing low price technically acceptable evaluation factors. The product is classified under NAICS 335999 and FSC 5999. The RFQ must be submitted via email to the buyer, Kim Vo, by 10 AM Pacific Time on September 09, 2024, with the FOB destination being the Puget Sound Naval Shipyard in Bremerton, WA. Offerors must complete the entire solicitation document, provide technical specifications and compliance certifications, and ensure current SAM registration. Any questions must be emailed to the buyer.

View the file

Other files for this federal contract opportunity

Other files attached to ACQUISITION RECORDER, newest first.
File Type Posted
SOLICITATION SPMYM2-24-Q-2480.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Main Unit specifications -42220109 Data Acquisition recorder

ITEM SPECIFICATIONS

Display 12.1-inch, TFT color LCD (1024×768) with capacitive touch panel

Channel

Module slot ... 9 slots Analog measurement ... Max 36 channels (when a 4-channel voltage module is used) Logic measurement ... Max 144 channels (when a 16-channel logic module is used)

Memory

Capacity 2GW (1GW/ ch during 2-channel measurement)

Division Memory

1, 2, 4, 8, 16, 32, 64, 128 division

Sampling Speed Memory recording ... Max 20MS / s (5ns) SSD recording ............. Max 1MS / s (1μs)

Storage Device

Solid State Drive (SSD) 256GB SD card slot with support for SD/ SDHC / SDXC. For data storage after recording.

USB memory using a USB port, for data storage after recording.

Printer

Printing Method

Thermal printing using a thermal head

Paper Width 219.5mm

Paper Feeding Speed

100mm / s to 1mm / min

Effective Recording Width

200mm・FS

Printer Functions

Memory Recording

Fast-speed event recording to memory

SSD

Recording

Recording of the input signal to the internal SSD

Printer Recording

Recording of the input signal directly to a thermal printer *Any combination of memory recording, SSD recording, and printer recording is possible.

Trigger

Uses Trigger for start recording operations, trigger for memory recording

Source Channels

Analog / logic signals, manual triggers and external signal triggers of each module.

Types Level triggered, window triggered (memory recording triggered ), bit pattern triggered

Trigger Filters Level hold, count

Trigger Output

Output signal when trigger conditions are met

Monitor

Y-T Waveform Monitor

Display time changes for a measurement signal in a pulse amplitude waveform

X-Y Waveform Monitor

Data is plotted with common points, with Signal 1 as the X axis and Signal 2 as the Y axis.

FFT Analysis Monitor

FFT analysis of the measurement signal of any two channels is performed, and the analysis results are displayed as the frequency axis data.

Recording and Display

Y-T Waveform

Displayed graphs ... 1 graph

Sheets ... Measurement channels are displayed in three sheets, and 48 channels are registered per sheet.

Grid ... 20 div vertically and 20 div horizontally

Time axis data count ... 100 data /div

Chart paper resolution ... Pulse amplitude axis 8dot/mm, time axis 80dot/mm (at 25mm / s)

X-Y Waveform

Data selection ... Select from input signal or recorded data

Displayed graphs ... 1 graph, 4 graphs

Grid ... 20 div vertically and 20 div horizontally

Sampling Speed ... Max 1KS / s

Locas ... ON/OFF of locas enabled (pen up& down)

Printing ... Print the plotted X-Y waveform with the printer

FFT Analysis

Data selection ... Select from input signal or recorded data

Sampling points ... 1,000, 2,000, 5,000, or 10,000 points

Max analysis frequency ... 1/2.5 times the sampling frequency

Displayed graphs ... 1 graph, 2 graphs. The Y-T waveform can also be displayed

Displayed graphs ... Waveform in time axis , linear spectrum, RMS spectrum, power spectrum, 1/1 octave, 1/3 octave,Transfer function, cross power spectrum, and coherence function

Window function ... Hanning window

Average processing ... Time axis simple addition average, frequency axis simple addition average,frequency axis exponentially weighted average, frequency axis peak hold

Interface

LAN

1000BASE-T (1Gbps) ... For control with communication command

COM

RS-232C ... For control by communication command or for UPS shutdown

USB Ver. 3.0 2 port ... For storage devices (USB memory)

Video Output DVI-D

Operating Environment Temperature 5 to 40°C Humidity 35 to 80% RH (without condensation)

Power Supply 100 to 240VAC, Frequency 50 to 60Hz Power consumption 300VA or less (under the maximum load conditions)

Dimensions 394 (W) × 336(H) × 200(D) mm *Excluding projections

Weight 9.5kg or less (main body only) 11 kg or less (when main body and four 2-channel voltage modules are installed)

Module Specifications 42220109-42220115

ITEM SPECIFICATIONS

•2ch Voltage Module- 4222110 item 2

Channel 2ch

Input Coupling AC, DC, and GND coupling

Maximum Sampling Speed

1MS / s

Frequency Response (- 3dB)

DC to 100kHz

A/D Resolution 16bit

Range Accuracy ±0.3% of RANGE *After offset cancellation

Allowable Input Voltage ±500V (DC or AC peak values)

Measurement Range 100mV to 500V (1-2-5 step)

Input Terminals and Models

Insulated BNC connector, insulated unbalanced input, isolation between each channel

Input Impedance 1MΩ

Low-Pass Filter 3, 30, 300, 3k, OFF

Anti-Aliasing Filter 10 to 40kHz

Maximum Rated Voltage To Earth

AC, DC300V

•4ch Voltage Module RA30-102

Channel 4ch

Input Coupling DC and GND coupling

Maximum Sampling Speed

1MS / s

Frequency Response (- 3dB)

DC to 100kHz

A/D Resolution 16bit

Range Accuracy ±0.2% of RANGE *After offset cancellation

Allowable Input Voltage ±200V (DC or AC peak values)

Measurement Range 1V to 200V (1-2-5 step)

Input Terminals and Models

Insulated BNC connector, insulated unbalanced input, isolation between each channel

Input Impedance 1MΩ

Low-Pass Filter 3, 30, 300, 3k, OFF

Maximum Rated Voltage To Earth

AC, DC300V

•2ch High-speed Voltage Module- 4220113 item 6

Channel 2ch

Input Coupling AC, DC, and GND coupling

Maximum Sampling Speed

20MS / s

Frequency Response (- 3dB)

DC to 5MHz

A/D Resolution 14bit

Range Accuracy ±0.5% of RANGE *After offset cancellation

Allowable Input Voltage ±500V (DC or AC peak values)

Measurement Range 100mV to 500V (1-2-5 step)

Input Terminals and Models

Insulated BNC connector, insulated unbalanced input, isolation between each channel

Input Impedance 1MΩ

Low-Pass Filter 5, 50k, 500k, OFF

Maximum Rated Voltage To Earth

AC, DC300V

•16-channel Logic Module

Bit 16bit

Maximum Sampling Speed

1MS / s

Input Type Voltage or contact, isolated input, no insulation between channels

Input Voltage Range 0~24V

Input Response Time 1μs

Input Terminal 8ch square connector x 2 ports

Notes Input by an optional IC clip or electrical clip code

•Remote Control Module

Features

Start/Stop by external signal, mark print, paper-feed, external sample

Input of external trigger signal, trigger signal output

Synchronous digital signal for AC distortion module

Module Name Specifications

2ch Voltage Module – 42220112 item 4

2ch, Sampling Speed 1MS / s, Maximum input voltage ±500V with anti-aliasing filter

4ch Voltage Module 4ch, Sampling Speed 1MS/ s, maximum input voltage ±200V

2ch High-speed Voltage Module 42220103 item

2ch, Sampling speed 20MS/ s, maximum input voltage ±500V

16-channel Logic Module 16 channels (voltage or contact)

Remote Control Module Remote control, TRIG IN/OUT, synchronization signal output when an AC distortion module is used

File details come from the government source that posted it. Updated .