This document describes the salient characteristics of a Handheld X-Ray Fluorescence (XRF) Spectrometer required for a federal contract opportunity. The key requirements include: dimensions, weight, excitation source, primary beam filtration, detector, power, display, temperature range, humidity, drop test, IP rating, pressure correction, operating system, USB connectivity, warranty, analysis capabilities, and accessories. The related federal contract opportunity is Solicitation Number SPMYM224Q2249 for a SPECTROMETER, issued by the Defense Logistics Agency Land and Maritime. The solicitation appears to be for the procurement of one Handheld XRF Spectrometer, along with a fan and portable workstation.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SPMYM2-24-Q-2249.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics for a Handheld X-Ray Fluorescence Spectrometer
19 July 24
1. HANDHELD X-RAY FLUORESCENCE (XRF) SPECTROMETER REQUIREMENTS:
a. One (1) Handheld XRF Spectrometer (V2CR-XXX-U, Vanta Core model handheld XRF analyzer, Evident Scientific Inc. Item # Q0204282or equal) to include the following minimum (or equal) features:
i. Dimensions (W x H x D): 10.4 cm × 29.6 cm × 24.1 cm (4.1 in. × 11.6 in. × 9.5 in.)
ii. Weight: 1.70 kg (3.75 lb) with battery, 1.48 kg (3.24 lb) without battery
iii. Excitation Source: Rhodium (Rh), Tungsten anode 8-40 kV Xray tube
iv. Primary Beam Filtration: 8-position auto selected filter per beam per mode; optional collimation to 3 mm diameter beam spot
v. Detector: Silicon drift detector
vi. Power: Removable 14.4 V Li-ion battery with hot-swap capability or 18 V power transformer 100–240 VAC, 50–60 Hz, 70 W max
vii. Display: 800 × 480 (WVGA) LCD with capacitive touch screen supporting gesture control
viii. Temperature range: −10 °C to 50 °C (14 °F to 122 °F), (continuous duty cycle with optional fan)
ix. Humidity: 10% to 90% relative humidity non-condensing
x. Drop Test: Military Standard 810-G 4-foot (1.3 M) drop test
xi. IP Rating and Detector Shutter: Dust protected and protected against water jets from all directions and solid detector shutter to help prevent detector damage
xii. Pressure Correction: Built-in barometer for automatic altitude and air density correction
xiii. Operating System: Linux cloud enabled with user fleet manager capability
xiv. USB: (2) USB 2.0 type A host ports for accessories such as wireless LAN, Bluetooth®, and USB flash drives, (1) USB 2.0 type mini-B port for connection to computer.
xv. Warranty: 3-year warranty
xvi. Primary Method: Alloy Plus Calibration for Vanta Core Series rhodium anode analyzers.
xvii. Analysis for metal samples measuring Mg, Al, Si, P, S, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, W, Hf, Ta, Re, Pb, Bi, Zr, Nb, Mo, Pd, Ag, Cd, Sn, Sb.
xviii. Accessories: Docking station, sealed rugged carrying case, 2 Li-Ion batteries, USB cable, USB storage with PC Software, 10 spare analysis windows, wrist strap and power supply, with USA power cord.
b. One (1) Fan (FAN-V, Evident Scientific Inc. Item # Q0200524 or equal)
i. Ordered with new instrument and factory installed. Fan ships from factory installed inside of XRF analyzer
c. One (1) Portable Workstation (WORKSTN-V-U, Evident Scientific Inc., Item # Q0200501 or equal)
i. Fully shielded and interlocked, rugged test stand for benchtop or remote controlled testing. Includes USA power cord and power adapter (70W, 18V, 3.9A).