This document outlines the salient characteristics for a dose calibrator and well counter procurement by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 21. The dose calibrator must include features such as a color touch screen over 8 inches, USB connections, printing capability, 28 programmable nuclide keys, a repository of at least 60 nuclides for selection, dose calibration and quality control functions, automatic chamber QC including linearity, a printer for records, a 256 channel MCA with spectral analysis, a well counter for wipe and laboratory tests, user-defined wipe test protocols, automated well QC including chi-square and MDA, manual and automatic ROI, a preset key for F-18, and printable reports. The device must also meet specifications for siting on a standard countertop, detector type, ion chamber components and functions, electrometer and linearity accuracy, power requirements, and compatibility standards.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C26123Q0972.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Item Specifications:
Essential/significant physical, functional, or performance characteristics.
· Device must include a color touch screen at least 8 inches in length
· Device must include USB connections
· Device must offer use printing capability
· Device must include at least 28 programable nuclide keys
· Device must offer a repository of at least 60 nuclides for the end user to choose from
· Device must include dose calibration, quality control and self-diagnostics
· Device must offer automatic chamber QC including linearity
· Device must include a printer for hard copy record data
· Device must include a 256 channel MCA with detailed spectrum for identification analysis
· Device must include a well counter for wipe tests and laboratory tests
· Device must include user-definable protocols for wipe testing
· Device must include automated well QC including Chi-Square and MDA
· Device must offer both manual and automatic ROI
· Device must include a pre-set key for F-18 measurements
· Device must offer printable reports for Well counter reports for autocalibration, system test, MDA, Chi-Square and laboratory tests
· Device must include automated linearity testing and geometry testing
· Device must be able to sit on a standard countertop
· Device well detector must be a Sodium Iodide (NaI) drilled-well crystal detector with programable warning triggers, and performs daily test, energy calibration and reproducibility tests
· Device ionization chamber must use a 12 atm Ultra-Pure Argon fill gas, detect activity up to 6 Ci, have a resolution of at least .001 MBq, have a response time of 20 seconds or less and perform daily Auto Zero, Background Adjustment, Voltage tests
· Test, Data Check, Accuracy and Constancy
· Device electrometer accuracy must be at least ± 2%
· Device linearity accuracy must be at least ± 2%
· Device must operate using standard 110V power
· Device must be plug and play
· Device must be able to detect activities as low as 1 nCi
a) Compatibility – N/A b) Industry Standard – yes c) Purpose – Provides accurate measurement of activity and preparation of doses for radiopharmaceuticals using in diagnostic imaging.
d) Governing Body or Organization – Veterans Health Administration Complete generic identification – Dose Calibrator and Well Counter Per VAAR 811.104 and 811.105, I certify the following regarding the above information:
· The salient characteristics listed above are essential to the Government’s requirements.