This statement of work specifies requirements for a laser induced breakdown spectroscopy analyzer for the Grand Junction VA. The analyzer must be able to detect carbon from 0.25-15% and silicon from 0-6.5%, have a spectrometer range of 190-420nm, weigh 3-6 pounds, be battery powered with a 2.5-3 inch display screen. It must be able to detect various elements, have 5+ years of warranty, and be delivered within the shortest time period. The related federal contract opportunity is solicitation number 36C25923Q0690 from the Department of Veterans Affairs for this LIBS analyzer, with delivery date being the primary evaluation criteria over price.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C25923Q0690.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
8/17/2021
Statement of Work LIBS Non-Destructive Testing Equipment
Grand Junction WCHCS 575
1. Purpose: To determine chemical composition of different alloys of piping, specifically cast Iron and forged carbon steel fittings. The Grand Junction VA requires a Laser Induced Breakdown Spectroscopy (LIBS) Gun.
2. Background: The Grand Junction VA requires equipment to determine chemical composition of cast iron and forged carbon steel in accordance with VHA operational alert and guidance document FY2022-01. This Guidance requires removal of all cast iron fittings to be replaced with carbon steel. Test methods to identify the difference between cast iron and steel are destructive methods that would render fitting inoperable for use. The WCHCS Engineering projects team determined that nondestructive LIBS spectrometry was an acceptable method to identify metals to be installed in the steam system.
3. Specifics: The instrument and its associated hardware for the Grand Junction WCHCS will have the following specifications:
A. LIBS Analyzer
a. Must use a Laser source to take samples chemical composition of metal
b. Must be able to detect Carbon within 0.25% to 15%
c. Must be able to detect Silicon to an accuracy within 0% to 6.5%
d. Must have spectrometer range from 190nm to 420nm.
e. Should be handheld device that weighs between 3 and 6 pounds.
f. Must be battery powered
g. Should have a display screen between 2.5 and 3 inches.
h. Must have library of grades of material to display once sample is taken
i. Must have ability to be calibrated.
j. Must not leave mark on sample material greater than 0.050 of an inch.
k. Must have argon shielding gas for laser.
8/17/2021
l. Must have range to detect elements C, Be, Mg, Al, Si, Ti, V, Cr, Mn, Fe, Co, Ni Cu, Cu, Zn, Zr, Pb, Bi, Ag, Se, Nb, Ta, Re, Mo, W.
4. Warranty: The Analyzer and associate hardware shall have a minimum warranty of 5 years.
5. Evaluation Criteria: The evaluation criteria are delivery date and price. Delivery date is more important than price. Delivery date is number of days from award to delivery to Grand Junction WCHCS
6. General:
The Analyzer shall be delivered to the Grand Junction WCHCS, 2121 North Ave, Grand Junction, CO 81501.
End of Statement of Work