This document contains the requirements for a hydrogen peroxide based low-temperature sterilization system to be procured through solicitation number 36C26223Q1546 by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22. The sterilization system must be manufactured by Advanced Sterilization Products Services, Inc. and meet numerous technical requirements including using gas plasma technology to break down hydrogen peroxide, limiting leak and gasket testing needs, providing cycle information and load placement graphics, notifications including automatic biological indicator reminders, and touch free cassette disposal. Additional requirements include endorsement from device manufacturers, hydrogen peroxide measurement for every cycle, an upgradable and expandable platform, minimum chamber size, removal of any old system, and included shipping and handling in the pricing. Power requirements and cord length are also specified.

View the file

Other files for this federal contract opportunity

Other files attached to 6515--Sterilizer, newest first.
File Type Posted
S02 - Attachment A - Sterilizer.xlsx XLSX spreadsheet
36C26223Q1546.docx DOCX document
S02 - RFQ - Sterilizer_RWH Edits.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment B Brand Name or Equal Description/Requirements:

Manufacturer: Advanced Sterilization Products Services, Inc.

Part Number: 10104-007-50 Nomenclature: Hydrogen Peroxide Based Low-Temperature Sterilization System Model Number: ST100NX ALLCLEAR 1-DR DUO Salient characteristics –

1. Low temperature sterilizers that use gas plasma technology.

2. Plasma breaks down hydrogen peroxide in the chamber, removes residual peroxide from the load, and helps make devices and trays safe to handle.

3. Limited/No weekly chamber leak test required.

4. Limited/No quarterly gasket inspection required.

5. Communication of information between Sterrad system, Velocity BI via ASP Access

6. Cycle information screens describing the types of devices that can be processed in each cycle.

7. Graphic display indicating proper load placement for the selected cycle.

8. On screen notifications, including automatic biological indicators reminders.

9. Touch free cassette disposal for staff safety.

10. Endorsement from device manufacturers.

11. H2O2 sterilant measurement accurately for every cycle via UV lamp.

12. Upgradable and expandable platform.

13. Chamber size no smaller than: 20.1”W x 28.9”D x 16.1”H

14. Removal of old System at no cost to the Government.

15. Shipping and Handling must be included in the item pricing.

16. Power Requirements:

· 208VAC, 60 Hz, 3-phase wye configuration circuit

· Separate neutral and ground conductors

· NEMA L21-30 5 wire grounding twist lock outlet

17. Power Cord Length: At least 9.8” (3 m)