S02 - RFI Attachment B.docx

DOCX document 19 KB Posted

Attached to
6515--INJECTION SYSTEM - RADIOLOGY Federal contract opportunity
Solicitation number
36C26223Q1458
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22

About this file

This document outlines the technical requirements for a contrast injector for mammography to be procured by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22. The contrast injector must be compatible with Hologic mammography units, support full DICOM compliance for contrast tracking, and interface with commercial PACS systems. Additional requirements include adjustable flow rates from 0.1 to 10 mL/sec in 0.1 mL increments, programmable volumes from 1 to 150/200 mL syringe capacity in 1 mL increments, and programmable pressure limits for various syringe sizes. The injector must support up to six phases, operate on 100-240 VAC power, and maintain syringe temperature at 35 degrees C +/-5 degrees. Installation is required along with delivery and includes all necessary equipment and cables. Shipping and handling costs must be included in item pricing and cannot be separately priced.

View the file

Other files for this federal contract opportunity

Other files attached to 6515--INJECTION SYSTEM - RADIOLOGY, newest first.
File Type Posted
S02 - RFI Attachment A.xlsx XLSX spreadsheet
36C26223Q1458.docx DOCX document
S02 - RFQ-36C26223Q1458_Radiology Instrument.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics: Contrast Injector for Mammography

· Must be compatible and suitable with contrast enhanced mammography.

· Must be compatible with Hologic mammography units.

· Must have interface for commercial Picture Archiving and Communications System (PACS).

· Must support full Digital Imaging and Communications in Medicine (DICOM) compliance for contrast information tracking.

· Must have floor/pedestal mount.

· Must have adjustable Flow Rate (range & increments):

· 0.1–10 mL/sec

· in 0.1 mL Increments

· Volume (range & increments):

· 1 mL to Syringe Capacity in

· 1 mL Increments

· Programmable Pressure Limit (psi/kPa):

· 150 mL and 200 mL Syringe: Choice of 50/345, 100/689, 150/1034,

· 200/1379, 225/1551, 250/1724, 300/2068, 325/2241

· Scan Reminders:

· 0–300 Seconds (5 minutes)

· in 1 Second Increments

· Pause:

· 1–900 Seconds (15 minutes)

· in 1 Second Increments

· Hold:

· Maximum HOLD Time is 20 Minutes

· Syringes (Volume capacity):

· 150 mL or 200 mL Sterile Disposable Syringe

· Maximum Number of Phases: 6

· Electrical Requirements:(VAC/Hz): 100–240 VAC, 50/60 Hz, 300 Volts-Amps
· Syringe Heat Maintainer:35 degrees C +/–5 degrees, 95 degrees F +/–9 degrees
· Head Mounting Options:Overhead Counterpoise System or Floor Pedestal (FloorPedestal includes an Integral IV Pole)
· DualFlow:Simultaneous Injection of Contrast and Saline is integrated inthe Certegra® Workstation (CWS) and included
· VirtualCare™ Remote Support*:Remote service on the injector and software to ensure faster uptime

Installation:

Along with delivery of radiology injection system, installation will be required and includes the injector head and pedestal, workstation monitor, all necessary cables, and Main Control Unit which is the computer itself. Basically, installation includes everything needed to allow the injector to work Shipping and Handling:

Shipping and Handling MUST be included in the item pricing and NOT SEPARATELY priced.