RT-PCR Specifics.docx

DOCX document 13 KB Posted

Attached to
CFX Opus 96 Real-Time PCR System Federal contract opportunity
Solicitation number
12405B23Q0083
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery

View the file

Other files for this federal contract opportunity

Other files attached to CFX Opus 96 Real-Time PCR System, newest first.
File Type Posted
Solicitation SAM.docx DOCX document
Clauses and Provisions.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for the Real-Time PCR System:

- Real-Time PCR System for gene expression analysis (relative quantity and normalized expression) for multiple genes, including PCR quantification with standard curve, end-point analysis and melting curve analysis.

- Flexible 96-well sample block for sample volumes ranging 10-50 µl.

- Optical Detection System with multiple filtered photodiodes and detection absorbance spectra in 450-684 nm/emission spectra in 515-730 range.

- Detection sensitivity: target gene equivalent to one copy of eukaryotic genomic DNA and fluorescin ≤ fmol.

- Multiplex analysis capable for up to 5 target genes per well.

- Reaction block temperature ranging 4-100 °C with average ramp rate of 3.3 °C/sec, including lid heating capable of 30-100 °C.

- Thermal accuracy within ±0.2 °C.

- Operational gradient ranging from 30 to 100 °C.

- Operating system using Windows 7 and above version (optimal Windows 10, 64-bit).

- Data analysis capable for graphic presentation, such as bar charts, dot plots, scatter plots, etc.) and basic statistical analysis, such as t-test and one-way ANOVA.

- Data export functions including graphs and spreadsheets compatible with Microsoft Excel, Word, or PowerPoint.