Tomey RT-7000 Auto Refractor/keratometer/topographer 12 December 2022

• Introduction: Auto-refraction is captured on every patient seen pre-operatively and post-operatively o This gives a starting point for the prescription to be treated by the surgeon o This gives how much residual prescription after surgery

• Keratometery is captured on every patient seen pre-operatively and post-operatively o Pre-Operatively we must make sure the patient’s cornea is not too steep

If the patient is too steep then the cornea may not flatten correctly o Post-Operatively we must make sure the patient’s cornea is not too flat

If the cornea is too flat is cause many issues:

• Glaucoma – loss of vision

• Prescription could be worse than pre-operatively

• Ectasia of cornea which is the thinning of the cornea which could lead to corneal perforation (rupture) If the cornea is too steep after surgery that can lead to many issues:

• Thickening of the cornea

• Loss of the endothelium layer cells o These cells do not grow back and can lead to haze o They can also lead to a corneal transplant

• Topography shows the valleys and peaks of the cornea o Topography shows where astigmatism is on the cornea o Topography shows where any residual astigmatism could be after surgery o Topography is done on all patients before surgery and after surgery

The topography is done at 12 post-op for extra help This shows if the astigmatism is with the rule or against the rule

• This machine will save us time and getting another machine for just topography

• Technical data of needed item:

o Various Corneal maps available for each patient

• Must be able to read sphere of +/- 25.00 diopters & +/- 10.00 diopters Cylinder

• The white to white that can be read on the above is needed to make sure the cone will fit for all

Lasik patients.

1. Background: The RT-7000 has been used by many of the technicians of Laser Eye. The RT-7000 cuts down optometrist and tech time capturing auto-refraction, keratometery, topography, ans white-to-white measurements. The RT-7000 is equipped with auto focus, the machine also can check tear film to prevent dry eyes, the many corneal maps that would be used by the surgeons, the machine can read +/- 25.00 diopters sphere and +/- 10.00 diopters cylinder.

2. Delivery Date: 14 days after receipt of order o Requested delivery dates: Any day but Wednesdays, that is surgery day o Base Access Point of Contact – o SrA Robin Claire Luna 937-257-8866 robinclaire.e.luna.mil@health.mil o SSgt Vivian Thompson 937-522-2756 vivian.e.thompson5.mil@health.mil

STATEMENT OF WORK

o SSgt Vivian Thompson 937-522-2756 vivian.e.thompson5.mil@health.mil o Surgical procedure dates – o Is hand delivery by sales representative required Y N N/A o Does sales representative need to be present during surgery Y N N/A

If yes please provide description of sales rep duties

3. Building address: 4881 Sugar Maple Dr. Bldg 830 WPAFB, OH 45433

4. Product Requirements:

o Size – The size is one size for all o Design – This item needs a 120VA plug.

o Performance Requirements – The RT-7000 Auto refractor/keratometer/topographer system is comprised of a 6.4 inch color LCD, easy & speedy touch screen alignment, diameter measurements of the cornea and pupil, Indices of KAI and KRI that show asymmetry and regularity of cornea, measures -25.00D to +22.00D sphere, 0D to +/- 10.00D cylinder, 5.00mm to 11.00mm corneal curvature measurement, 0.2 seconds refraction measurement time, 0.1 seconds keratometer measurement time, minimum measurement refraction steps: 0.01D sphere, 0.01D cylinder, 1 degree axis. Minimum measurement corneal curvature measurements: 0.01mm.

o Accessories – The only accessory needed is a printer o Configuration – There is no configuration needed.

o Installation & Testing – The item will need to be installed and tested by MEMO o Training – The training will be on the job and will take less than a day.

o Warranty – The manufacturer warranty is 1 year from date of purchase.

o Standards, guidelines or approvals – The item will be watched over by MEMO

Check Box3: Yes
Check Box4: Yes