RFQ-OH-15-00165
Attachment A: Specifications Autoanalyzer and associated equipment
The contractor shall provide an autoanalyzer and associated equipment as specified below:
· Flow injection autoanalyzer and integrated ion chromatograph capable of analyzing fresh, brackish and saltwater samples
· Block sample digester (holding up to at least 40 samples at a time)
· Computer with Windows 7 operating system, monitor and associated software to interface with auto analyzer and accessories
· Heater apparatus required for analysis of certain analytes below
· Detector assembly
· Auto sampler capable of holding up to 360 samples at once
· At least five (5) sample channels
· At least a sixteen (16) channel reagent pump
· Integrated sample-dilution system
· Equipment/Manifolds to enable the analysis of the following:
| Analyte: | Detection range: | |||
| · Alkalinity | at least 10 – 200 mg CaCO3/L or better | |||
| · Ammonia | at least 0.03 – 3 mg N/L as NH3 or better | |||
| · Total hardness | at least 10 – 200 mg CaCO3/L or better | |||
| · Total Kjeldahl nitrogen | at least 0.25 – 5 mg N/L or better | |||
| · Total Kjeldahl phosphorus | at least 0.25 – 5 mg P/L or better | |||
| · Nitrate + nitrite | at least 0.005 – 0.1 mg N/L or better | |||
| · Orthophosphate | at least 2 – 80 µg P/L as PO42- or better | |||
| · Reducing sugars | at least 20 – 200 mg glucose/L or better | |||
| · Silicate | at least 10 – 80 µg SiO2/L or better | |||
| · Urea | at least 0.2 – 20 mg N/L as Urea or better |
· Anions:
| Anion: | Lower Detection Limit: | |
| · Bromide | at least 0.020 mg Br-/L or better | |
| · Chloride | at least 0.008 mg Cl-/L or better | |
| · Fluoride | at least 0.008 mg F-/L or better | |
| · Nitrate | at least 0.008 mg NO3-/L – N/L or better | |
| · Nitrite | at least 0.008 mg NO2-/L – N/L or better | |
| · Phosphorus | at least 0.015 mg HPO42- – P/L or better | |
| · Sulfate | at least 0.015 mg SO42-/L or better |
· On-site equipment installation and demonstrated proof of performance
· Training of equipment and software setup and use
· The computer equipment supplied by the contractor shall be able to accept the EPA image for Windows
· The equipment supplied by the contractor shall be UL (Underwriters Labs) listed or a recognized 3rd party listed. If not normally listed, the contractor shall ensure field listing be done at the manufacturer's location and paid for by the manufacturer.
Delivery: FOB Destination, 60 days ARO