AUTOCLAVE REQUIREMENTS
(Equal to BMT Sterivap Steam Sterilizer: Model SP6612-1)
· Double Microprocessor controls with double instrumentation for maximum load safety
· 8.2” Color touch screen operator interface
· 316L Stainless steel chamber and 316Ti jacket with 15 year non-prorated warranty
· Hinged fascia panels with key lock for ease in maintenance access
· All stainless steel frame – needed for corrosion resistance
· Mechanical vacuum pump for consistent vacuum performance
· Water conservation for up to 75% reduced water consumption
· Space saving design-less floor space required
· Automatic Motor driven precision sliding doors – vertical or horizontal – not swinging.
· High grade non-proprietary components
· Double pressure sensors ensure the no pressure inside the chamber before unlocking of doors can occur.
· Custom manufactured so that all serviceable parts are on the right side of the unit
· Stainless steel chamber piping
· Pneumatic valves constructed from brass
· Up to 7 validated preprogramed cycles with at least these cycles: Dry Goods Cycle with Vacuum drying; Dry Goods Cycle with Pulsed Air/Vacuum drying; Solution cycle with gravity air removal/ramped exhaust; Solution cycle with air over-pressure/jacket water cooling.
· Up to 20 total cycles available with unlimited cycle storage utilizing the chip card system
· 4.5” wide printer with digital and graphic print out
· Dual temperature sensors to provide precise control, evaluation and regulation of temperature inside the chamber and to ensure sterility has been achieved.
· Dual absolute pressure sensors for precise control of pressure and vacuum inside the chamber and pressure inside the jacket and to ensure sterility has been achieved.
· Optional load probe for solution cycles
· Internal dimensions minimum: 25” (w) x 27” (h) x 52” (d)
· External dimensions maximum: 52” (w) x 75 (h) x 63” (d) – must fit into a pre-existing room with no modifications.
· Chamber volume minimum: 21.5 cubic feet (610 liters)
· Excellent customer service and customer support
· Field serviceable by experienced factory trained engineers
· ASME safety relief valves
· One 1 1/2” validation port with sealing gland for up to 12 probes
· Auto-diagnostic system on OIT with printout
· Wall finishing frame for recessed application
· Auto Start-up and Shut-down
· Loading Cart with two shelves
· Transfer carriages
· 208V three phase 60 Hz power supply
· Shipping
· Training/installation/warranty service by local authorized service group