Revised_Final Specs for FIA.docx

DOCX document 17 KB Posted

Attached to
Flow Injection Analyzer, 3-channel, Brand Name or Equal Federal contract opportunity
Solicitation number
12905B22Q0080
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area

View the file

Other files for this federal contract opportunity

Other files attached to Flow Injection Analyzer, 3-channel, Brand Name or Equal, newest first.
File Type Posted
Solicitation clauses-provisions-TandCs.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

The USDA, ARS, Northwest Irrigation and Soils Research Unit, Kimberly, ID requires the following specified components for a 3-Channel Flow Injection Analyzer.

Specifications for a Brand Name or Equal to FIAlab Instruments FIALYZER 3000 3-Channel Flow Injection Analyzer in accordance with FAR 8.405-6 Limited Sources:

1. The detector must be an array spectrometer. Displays the full absorbance spectrum in real time.

2. The system must have an EPA approved fluorescence method for TKN & Ammonia without the need to switchover manifolds.

3. The footprint must be no more than 30 inches deep and 30 inches wide to fit on available lab bench.

4. The system must utilize a 1.7 mm ID bore inside Lab on Valve to prevent clogging instead of the 0.8 mm ID (or smaller) typically used in previous Flow Injection systems.

5. Throughput for Ammonia (Salicylate) must be 80 samples/hour.

6. Full view of fluid movement in the manifolds is required.

7. The system must include analyzer accessory with absorbance flow cells that provide an up to 100 mm path.

8. The system must include an FIA instrument that is fully redundant with each channel. Each channel must have its own24V power supply, LED light source, and detector. The unit must operate from a standard electrical outlet, 240v max.

9. Must be able to analyze samples while running with different variables, i.e., changing wavelengths.

10. The software must be compatible with Windows 10 or Windows 11 and written in C++. There should be no restriction on the number of users or license use.

11. The system must be compatible with and include an Autosampler that holds up to 240 samples in 15ml centrifuge tubes and include four racks.

12. The software must include features such as automated QC/QA protocol to monitor accuracy and precision along with QC limits for each QC check standard (relative and absolute).

13. Will include on-site installation and training for two or more users.

14. The system must meet “Buy America” statute.

15. The quote must include a minimum two-year non-consumable parts and service warrantee on the machine.