GENERAL SPECIFICATIONS

1. Four-channel automated flow injection analysis system set up to run four or more colorimetric assays, four at a time.

2. Control software package included.

3. Installation and 2-day training included.

4. Flow injection module for nitrate

5. Flow injection module for nitrite

6. Flow injection module for ammonia and total Kjeldahl nitrogen (TKN)

7. Flow injection module for phosphate

8. Flow through heater

9. Eight channel extended pump

10. Autodiluter

11. Extended service plan (2 years past manufacturer warranty)

a. Extension of warranty on non-consumable parts from 1 year to 2 years

b. One service visit

c. Phone, email and remote desktop support

d. Free software upgrades REQUIRED Specifications required for this purchase:

Light Detector: Array spectrometer with wavelength selection made simply from the software. REQUIRED to display the full absorbance spectrum in real time.

Speed. Throughput ability for Nitrite - 220 samples/hour required Software. Software must be compatible with Window10 and written in C++. Must be able to analyze samples while running with different variables—“analyze on the fly” Valve Stator. Full view of fluid movement in manifold REQUIRED.

Small analyzer footprint. Lab desk space is limited; small analyzer footprint required.

Long absorbance flow cells. Analyzer accessory must include with absorbance flow cells providing an up to 100 mm flow path.