This sources sought notice requests capability statements for a robotic single-cell printer with washing station. The National Institute on Aging requires equipment to efficiently separate single cells and place them into 96- and 384-well formats for single-cell omics, low-volume assays, and cell line production. Interested small businesses should submit documentation demonstrating ability to meet specifications including a F.Sight Omics cell printer with 150 pL droplet volume and I.Dot DW non-contact low volume dispenser. Capability statements are due by June 21st and should include company information, technical qualifications, and confirmation the required equipment can be delivered within 30-60 days of order receipt. The National Institutes of Health National Institute on Drug Abuse will evaluate responses to determine an appropriate acquisition method for this requirement.

View the file

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PURCHASE DESCRIPTION

1. Statement of Need and Purpose: NIA utilizes single cell omics/cell editing techniques (e.i. CRISPR), and requires equipment to efficiently separate out single cells, and placed it into into 96- and 384-well format.

2. Background Information and Objective: Single-cell omics, Low-volume assays (WGA, PCR, NGS setups) and cell line production are fast developing scientific areas. Single cell omics required appropriate instruments for this work and suggested equipmets will serve as approriate tool.Currently NIA has a QPCR machine where we can analyze 384-well plates. The process of aliquoting reagents by hands to 384-wells currently take about 1.5 hour with reduced reproducibility. The objective is obtaining a Robotic cell sorting/ washing station with the ability to make 1 plate in 25 seconds with very low error. Combination of these two instruments will give the National Institute on Aging (NIA) a huge increase in productivity, when NIA can separate and analyze 384 single cells within several minutes.

3. Generic Name of Product: Robotic single-cell printer with washing station, for handling liquid cell suspension for WGA,PCR, NGS preparation..

4. Purchase Description: F.SIGHT OMICS (single cell priner); I.DOT DW (Non-contact, low volume dispensing system), C.WASH (centrifugal plate washer). Company Dispendix (CELLINK LLC)

3. Salient characteristics: NIA requires a robotoc single-cell printer with washing station with the following characteristics:

F.SIGHT OMICS (Cell Printer):

Feature
Specification
Dimensions (W x D x H)
600x400x240 (mm)
Weight
35kg
Power sockets required
2x
Input Voltage [VAC]
100-240
Input voltage frequency [Hz]
50/60
Max. input current [A]
1.4
Max. power consumption [W]
135
Input fuses
2x 3.15 A quick react
Internal voltage levels [VDC]
5, 12, 24
LED source 1 (DQC)
white 1 W
LED source 2 (Detection)
red 3.0 W
Embedded Laser source 1
473 ±1 nm 100 mW ±5 %
Embedded Laser source 2
488 ±1 nm 100 mW ±5 %
Embedded Laser source 3
561 ±1 nm 100 mW ±5 %
CE mark
yes
CB mark
yes
QPS mark
yes
RoHS compliant
yes
WEEE
yes
Laser protection class DIN EN 60825-1: 2015
Class 1

FDA 21 CFR 1040.10 & 1040.1

Droplet volume
150 pl
Dead volume
< 1 ul
microfluidic outlet (nozzle)
40 x 40 ìm
Included warranty
12 months

I.DOT DW (Non-contact, low volume dispensing system):

Feature Specification
471 mm x 575 mm x 378 mm
Dimensions (L x W x H)
AC 100-120 V, 50/60 Hz, 10A or AC 200-240 V,
Power supply
AC 100-120 V, 50/60 Hz, 10A or AC 200-240 V,

50/60 Hz, 5A

Compressed air supply
Filtered, oil-free, dry air, 6–10 Bar (8–145 PSI)
Weight
48 kg (106 lb)
Destination
All SBS-format microplates not exceeding 127 mm

x 85 mm x 54 mm (L x W x H)

Dead volume
<1 uL
Dispensing volume per well
Standard plate: 8 nL to 80,000 nL (80 μL)

Deep reservoir plate: 8 nL to 500,000 nL (500 μL)

Volume unit step
0.1 nL
Dispensed volume CV
<5% (for >100 nL)

<8% (for <100 nL)

Liquid viscosity range
1 to 100 cP (mPa-S)
Nozzle diameter
60 um to 200 um
Source fluid temperature range
4° C to 37° C

(requires the Cooled Source Tray for low temperature)

Temperature accuracy
±0.5° C
Target plate temperature range
Room temperature

C.WASH: C.WASH centrifugal plate washer. Includes instrument with pump for waste removal, touch pad, user-friendly software, free software updates for 12 months, 4 liquid input channels, 2 standard carriers for 96/384-well plates, power cables, and 2 500 ml glass bottles for media, angled 8-pin and 16-pin dispenser for 96/384-well plates, 1x 500 ml C.WASH cleaning reagent.

Quantity: F.SIGHT OMICS (single cell priner)- 1; I.DOT DW (Non-contact, low volume dispensing system) - 1, C.WASH (centrifugal plate washer) – 1.

4. Delivery Date: 30-60 days after receipt of order Version 11-30-2016-1