Purchase Description-75N95023Q00442.pdf
PDF 236 KB Posted
- Attached to
- Automated liquid dispensing device Federal contract opportunity
- Solicitation number
- 75N95023Q00442
About this file
This purchase description outlines requirements for an automated liquid dispensing device to support research on Alzheimer's disease and related dementias. The National Institutes of Health National Institute on Drug Abuse seeks to procure a Mantis V3.3 instrument with integrated RFID and chip changer capacity of six chips. Accompanying items include a continuous flow option, base unit, remote user training, silicone chips in varying volumes, a PCR plate adapter, nozzle kit, input tube holder, locating plate, continuous flow chips, and CF cap kits. The instrument must have a small footprint and capabilities such as non-contact microplate liquid delivery in specified volume ranges and well configurations, fast high-volume dispensing with continuous flow, and compatibility with reagents, cells, and beads. Delivery within three months of order is required, along with a one-year manufacturer warranty effective upon receipt.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Invoicing Instructions With IPP.pdf | ||
| FAR 52.212-5 Jun 2023.pdf | ||
| FAR 52.204-26 Oct 2020.pdf | ||
| FAR 52.204-24 Nov 2021.pdf | ||
| FAR 52.212-4 Addendum.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Version 11-30-2016-1 FAR Part 11—Describing Agency Needs
PURCHASE DESCRIPTION
1. Statement of Need and Purpose: This acquisition is necessary for automation of the culture techniques used in the iPSC Neurodegenerative Diseases Initiative (INDI) Project within the new Center for Alzheimer’s Disease and Related Dementias.
2. Background Information and Objective: imaging technology improves to capture The Government is working to create a national resource of isogenic sets of Alzheimer’s Disease and Related Dementias mutated iPSC lines, complimented by foundational phenotypic datasets from differentiated disease-relevant cells derived from these lines.
This resource will substantially accelerate the pace of mechanistic discovery regarding how ADRD mutations cause disease. An automated liquid dispensing device is needed to create a high through put workflow for processing tissue culture cells to enable genetically engineered iPSCs.
3. Generic Name of Product: Automated liquid dispensing device
4. Purchase Description:
Description: Quantity Cat no:
Mantis V3.3 with Integrated RFID, Chip Changer - 6 Chip capacity and Mantis Controlling Software Instrument License
1.00 233095
Mantis Continuous Flow Option 1.00 MCF
One LC3 Base Unit including RFID Reader 1.00 290595_A
Mantis Remote User Training 1.00 360787
Mantis Chip - Silicone, HV (1uL & 5uL), RFID, PI. Pack of 12 1.00 233648
Mantis Chip - Silicone, LV (0.1uL & 0.5uL), RFID, PI. Pack of 6 1.00 233581
PCR Plate Adapter with wide conical pins for automated plate handling 1.00 232400
Mantis Nozzle Kit - Silicone (Includes 0.1, 0.2, 0.3, 0.4, 0.6mm ID, 2 each) 1.00 233036
Mantis Silicone Input Tube Holder for 15mL or 50mL source tubes 1.00 233858B
One Right Side LC3 Locating Plate for use in the set up of one LC3 unit attached to a Mantis. Includes a waste station adapter.
1.00 290604
Mantis Chip - Silicone, Continuous Flow (CF), RFID, PI. Pack of 6 1.00 233127 Mantis CF Cap - Nalgene Bottle, Silicone Harness, pack of 6. 1.00 233125
Shipping 1.00 NA
Salient characteristics All deliverables required under this contract shall be packaged, marked and shipped in accordance with Government specifications. At a minimum, all deliverables shall be marked with the contract number and Contractor name. The Contractor shall guarantee that all required materials shall be delivered in immediate
Version 11-30-2016-1 FAR Part 11—Describing Agency Needs usable and acceptable condition.
Vendor must be able to provide these specifications:
Instrument must have a footprint no larger than (6.9” x 8.5” x 9.3”)
Instrument must include a microdiaphragm pump technology to provide non-contact delivery of desired volumes of liquids into wells of SBS based 96, 384 and 1536 well microplates, as well as deep well blocks and vials.
Instrument must allow for low volume dispensing from 100 nL and up
Instrument must allow for low dead volume at 6uL
Instrument must allow for fast dispensing of high volumes with a continuous flow option.
Instrument must allow for dispensing reagents, cells (mammalian and bacterial) and beads (magnetic and non-magnetic)
5. Delivery Date: Delivery date of 3 months after receipt of order (ARO).
6. Period of Performance: Manufacturer warranty for one calendar year beginning date of receipt.