X-Ray Computed Tomography System

Summary This procurement is for a turn-key, state-of-the-art, micro-focus X-ray computed tomography (CT) system. The system must be able to scan parts up to 18” in width and 24” in length. The system must be a fully shielded and interlocked cabinet-style unit. It must comply with all applicable industry regulations. The system must include a flat panel detector, microfocus X-ray Tube, digital radiography (DR) image processing acquisition workstation and software, and a separate CT reconstruction workstation including software. The system and components must meet the minimum technical requirements listed below.

System Requirements

· 2D X-ray inspection

· 2D CT Slice reconstruction

· CT slice acquisition and volume reconstruction for 3D inspection

· 3D internal and external surface scanning

· Automated spiral acquisition and reconstruction using a digital flat-panel detector. Technique used for tall objects that may exceed the detector height or for reduction/elimination of cone-angle artifacts.

· Automated panel/specimen shifting that allows image stitching to achieve a larger field of view than the detector panel size and results in one seamless image.

· Ability to achieve enhanced resolutions above the native panel resolution. Analysis routines will be used to increase resolution in conjunction with slight movements of the manipulator.

· Time and motion analysis. The system shall be capable of creating and analyzing moving or changing parts as dynamic, time-based data sets. The system shall be able to acquire time-based data in either a continuous or stepped mode. Continuous scan mode refers to repeatedly scanning the part while motion or change is occurring at a speed that is slow relative to the time it takes to complete a full scan. Stepped scanning refers to re-scanning the part multiple times after the part is subjected to some condition that will cause motion or a change in the part. The system shall be able to generate time-sequenced images (videos) showing the motion or change in the part resulting from the continuous or stepped scans.

· Measurement tools for distances, polyline lengths, and angles.

· Automated or efficient process (<15 minutes) for calibration and alignment for CT reconstruction.

System Components (minimum requirements)

Cabinet:

· Radiation shielding to energies of 225kV constant potential. The system shall meet the requirements of 21 CFR 1020.40 Performance Standards for Ionizing Radiation Emitting Products, Cabinet X-ray Systems.

· The system shall have primary and secondary fail-safe interlocks.

· The system shall have a warning light to indicate when X-rays are being produced.

· The system shall have interior lighting.

· The system shall have a lead-glass viewing window.

Motorized Manipulator System:

The manipulator system allows for motorized control of the positioning device and associated X-ray system.

Required Features:

· All drives are variable-speed with independent axis control, allowing simultaneous movement of all axes.

· Desktop control console (software or hardware) to control manipulator system.

· Focal Distance: Variable; up to 52" (132cm).

· Scanning envelope: parts up to 18” diameter and 24” in length

· Part travel:

o Vertical - 24” (60cm) o Horizontal - 15” (38cm) standard, 21” (53cm) for specimen loading o Z axis - 52” (132cm) o Rotation - 360 Degrees. Micro CT precision stage.

· Motorized detector mount with translation to adjust the focal distance and lateral shifts of 32”

· 50lb (23kg) stage capacity for both DR and CT.

Detector Lateral Shift Lateral shift capability for enhanced image quality or image stitching capabilities for larger specimens.

Dual Head X-ray tube:

· Demountable dual head tube (Exchangeable transmission and directional tube heads)

· High voltage generator

· Vacuum system

· Cable and hose set

· Spare parts / service kit

· Directional head configuration

- Micro-focus mode

- High-power mode

- 20-225kV, 50-3000uA, up to 320W tube power

- Closed Loop Water Cooler

· Transmission head configuration

- Nanofocus mode, detail detectability <0.5 micron

- Microfocus mode

- High-power mode

- Max 80W tube power

X-ray Control Software:

· Windows based

· Automatic tube warm-up

· Automatic filament current adjustment

· Automatic focusing and electron beam alignment

· Intensity control

X-ray detector:

· Sensor Area: 16” x 16” (410mm x 410mm)

· Premium CT Grade (best quality)

· Cesium Scintillator (best contrast)

· 200 micron pixel pitch. 2048 x 2048 pixels

· 16-bit resolution

· Shielded to 225kV

· Premium CT Grade (best quality)

Computer hardware and Software:

The system shall include two computer workstations. The Digital Radiography (DR) Workstation shall be used to operate the system, acquire data, and for DR. The CT Reconstruction workstation shall be used for reconstructing and viewing 3D image data.

DR Workstation:

· Windows® 10 Operating System

· Eight-Core Intel® Xeon® Processor

· 8 GB RAM

· 1 TB SSD Operating System Drive

· 3TB of Data Storage in RAID 0 Configuration

· DVD+/-RW Drive

· 10/100/1000 Network Interface

· NVidia High Performance Display Adaptor

· 30” (762mm) 4 Mega Pixel Color Display

· Keyboard and Mouse

DR Software:

High-performance acquisition software with a comprehensive set of image analysis tools.

· Windows® based

· Non-proprietary image storage format.

· High performance image processing and measurement functions

· Live imaging and averaging

· Live measurements with 1 pixel resolution

· Live image gain/offset calibration and defective pixel correction

· Video capture into AVI files.

· Support for digital flat panel detectors and linear diode arrays (LDAs) at 8, 10, 12 and 16 bits.

· Read and store images in TIFF 32 bit / 16 bit / 8 bit, BMP, JPEG, and RAW formats

DR CT Acquisition Software:

The CT acquisition module will provide automated CT acquisition capabilities.

· Capable of CT acquisition up to 10,000 angle steps.

· Operator record.

· Fast scan option for continuous CT acquisition of projections (ability to acquire thousands of projections in less than 10 minutes).

· Automated spiral acquisition mode (coordinated manipulator and image acquisition for tall specimens).

· Automated panel/specimen shifting and acquisition that allows image stitching to achieve a larger field of view than the detector panel and results in one seamless image.

· Acquisition mode for enhanced resolutions above the native panel resolution.

· Acquisition for time and motion analysis. The system shall be able to acquire time-based data in either a continuous or stepped mode. Continuous scan mode refers to repeatedly scanning the part while motion or change is occurring at a speed that is slow relative to the time it takes to complete a full scan. Stepped scanning refers to re-scanning the part multiple times after the part is subjected to some condition that will cause motion or a change in the part.

CT Reconstruction Workstation:

· Windows® 10 Operating System

· Dual Six-Core Processors (Total 12 Cores)

· 256GB RAM

· 1 TB SSD Operating System Drive

· 24 TB Hard Drive with High Speed Raid 5 capabilities

· 4 NVidia High Performance Display Adaptor for Display and Reconstruction

- 96GB GPU Memory

- 15,360 Processing Cores

- 3D Rendering capabilities

· DVD+/-RW Drive

· 10/100/1000 Network Interface

· 30” (76cm) high resolution flat panel monitor

· Keyboard and mouse

Computed Tomography (CT) Software:

CT software program that includes CT geometry calculation, cone-beam reconstruction, fan-beam reconstruction, 3D real time visualization and analysis tools.

· Software is Windows based and compatible with Windows 7 or higher.

· Process that steps from creation of the project to the final results that includes calibration and controls for the reconstruction process.

· Cone-Beam Reconstruction – GPU-based processing for fast reconstruction of CT data sets

· Fan-Beam Reconstruction – GPU-based processing for fast reconstruction of fan beam sinograms

· Real-time volume rendering with multi-plane slicing, density segmentation, and pseudo color capabilities with zoom, pan, and rotation.

· Export of slices at any angle in the X, Y, or Z plane

· Surface extraction to CAD compatible format: Export to STL (Stereolithography), cloud points, OBJ, DXF, WRL, PLY.

· Spiral CT reconstruction for scanning tall objects or areas with reduced noise.

· Software should include ring artifact reduction tools using a combination of hardware and software to reduce or eliminate ring artifacts.

· 3D reconstruction of stitched images to form one seamless image volume

· 3D reconstruction mode to achieve enhanced resolution from panel shifted images

· Time and motion analysis (see System Requirements)

Miscellaneous

Copper Filter Kit:

Set of copper filters for use with microfocus tubes:

· Selection of 0.005", 0.010", 0.020", 0.040", 0.060", 0.080" thick filters

Installation and Training

DR / CT System Training: Multi-day on-Site DR/CT system operation training. Training shall include DR, CT, operation, maintenance, and safety of the delivered CT system. This training shall be provided as part of the system delivery and acceptance

Digital Radiography & CT Training (40-hour minimum) 2 Seats 40 hour ASTM/ASNT Certified DR/CT training course.

This course should follow the recommended curriculum of FWG-IDR, MAI, ASTM and ASME in a format for compliance with NAS 410 and SNT-TC-1A type Qualification/Certification programs.

Advanced CT Training (1 Seat) Advanced CT Training (4-day minimum) that addresses advanced capabilities of the Industrial CT technology and includes CT Technique Development and CT Reconstruction. The class shall focus on the advanced capabilities of the delivered system and Industrial CT technology.

Maintenance

12 Month Hardware Warranty with one preventative maintenance visit.

First Year DR & CT Software Maintenance/Updates Local support by main manufacturing facility

Technical Manual The system shall include a user’s guide technical manual that covers system operation, routine maintenance and basic troubleshooting information. The standard commercial user/technical manual(s) for the hardware and software should meet this requirement.

Delivery Shipping Included FOB DESTINATION The equipment shall be delivered and installed within 20 weeks of receipt of the contract.