X-Ray Vault Requirements
· Dimensions: Minimum 8ft x 8ft x 8ft
· Capable of containing minimum 160kV
· Cable baffles for x-ray unit controls and auxiliary equipment
· Interior fluorescent lights with switch
· 120V, 15AMP outlet per wall
· Interlock and indicator system complying with 21 CFR 1020.4
· Electrical schematics
· Assembly drawings of mechanical systems
· Bill of Materials
· Operator’s & Maintenance Manuals
· Comply with Title 21 CFR Chapter I Subchapter J Sec. 1020.40 requirements
· http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?FR=1020.40 Optional
· Motorized sliding door
· Small overhead hoist for mounting tube head