Specifications for the Liquid Handling Platform:
· be Illumina-validated library preparation protocols, including Nextera DNA Flex, rapid exome capture, TruSeq protocols, and DNA Nano protocols;
· be capable of SPRI bead purification and normalization (including integrated 96-well magnetic plate)
· Elute in volume less than 10 ul;
· be capable of pipetting down to 1 ul in multiple channel mode;
· allow for scalable tip loading, including single tip loading, to minimize waste;
· have flexibility to accommodate 1 to 48 libraries per run;
· have a heated mixing platform;
· fit on to a lab bench space 30 in. x 36 in;
· include unit enclosure for sample protection;
· brief user pre-run preparation time;
· include instrument set-up and calibration;
· include user platform training;
· include user method training;
· be adaptable to custom protocols