NIST HARDWARE SPECIFICATION

EQUIPMENT:

A cellular wireless device test system consisting of a reverberation chamber, base-station emulator, channel emulator, and rotating platform.

Background:

NIST is supporting the development of test methods, best practice guidelines, and uncertainties for the cellular wireless community. New over-the-air test methods for cellular wireless devices that utilize reverberation chambers are under development. NIST requires hardware that will allow development and verification of proposed methods.

Objectives:

The objective is to procure instrumentation that will enable NIST to replicate proposed test methods and develop new or refined test methods for the cellular wireless industry. Both single-input, single-output (SISO) and 2 x 2 multiple-input, multiple-output (MIMO) devices shall be tested. Handheld devices shall be tested in the new chamber, while large-form-factor devices shall be tested in existing NIST chambers. Consequently, NIST requires four line items: (1) A reverberation chamber test chamber set-up for handheld devices; (2) a base-station emulator; (3) a channel emulator; and (4) a rotating turntable for testing large-form-factor devices. The first three line items shall be capable of measuring both SISO and 2 x 2 MIMO test conditions. Five years of annual calibration of the base station emulator and channel emulator is required. A warranty period of a minimum of 3 years for these units is required.

Delivery:

Delivery of all components shall be less than 12 weeks.

Line Item 1: Reverberation Chamber Test System

Parameter
Conditions
Min
Typ
Max
Units
Frequency range of operation
Frequency range may exceed typical
--
0.7-6
--
GHz
Shielding performance of chamber from ambient
To be measured and certified by manufacturer for at least the minimum and maximum frequencies of operation
100
120
--
dB
Accuracy
Accuracy in reference power transfer function when averaged over complete stirring sequence, 2 x 2 MIMO mode
--
0.3
0.5
dB
Mechanical mode-stirring paddle
Mechanical paddle may rotate or move laterally
1
2
--
each
Rotating platform
Supports loads up to 20 kg
1
1

each

Stirring sequence control
User must be able to specify and automatically control all stirring mechanisms in both stepped and continuous stirring modes, with at least the specified number of stirring states (excluding frequency stirring)
1000
10,000
--
samples
Interior side dimension
Interior dimension on each side must be at least 3 at 700 MHz (=~0.428m)
1.28
--
2.5
m
Working volume dimensions
Defined as volume where device under test may be placed such that it may maintain a distance from any metallic or RF absorbing surface of at least /2 at the lowest frequency of operation (~0.214m). May be cylindrical. Minimum dimension for any single side or height of working volume is 0.5 m.
0.5
1
6
m3
Antennas
Two omnidirectional reference (transmit) antennas, two directional measurement (receive) antennas required. Must operate over “frequency range of operation” stated above.
4
6
--
each
Minimum root-mean-square delay spread (RMSDS)
Sufficient RF absorber shall be provided to produce an RMSDS of 50 ns (operation shall typically utilize longer RMSDS) without diminishing working volume.
50
50
--
ns
Bulkhead RF feedthrough connectors
SMA (female-female) or Type N (female-female) is acceptable. Must operate over “frequency range of operation” above and be compatible with 2x2 MIMO operation.
6
8
--
each
Bulkhead control feedthrough connectors
USB (female-female)
1
2
--
each
Bulkhead feedthrough panels
Blank panel or pipe with caps to allow outside-to-inside access. Shielded when not in use.
2
4
--
each
Door to chamber
Hinged door with user-operated latch
1
1
--
door with user-operated latch
Operating Ambient Air Temp
Unit and electronics shall operate over a standard commercial operating range
--
Commercial
--
--

Line Item 2: Channel Emulator

Parameter
Conditions
Min
Typ
Max
Units
Frequency range of operation
Frequency range may exceed typical
--
0.7-6
--
GHz
RF channel bandwidth
Instantaneous bandwidth
40
40
--
MHz
RF input power range
Ratio of minimum to maximum input power with SNR > 35 dB
50
55
--
dB
RF output level resolution
Over entire input power range
--
0.1
0.1
dB
Error vector magnitude
Root-mean-square EVM_RMS for WCDMA (3.84 MHz BW)
--
-50
-45
dBm
Number of channels
Capable of supporting 4G LTE with 2x2 MIMO
2
2
--
channels

Taps per channel

24
48
--
taps
Channel models
At a minimum: (1) 3GPP/CTIA spatial channel models for MIMO simulations, (2) user-defined channel models
2
--
--
channel model modes
Doppler
Emulation of Doppler shift
2
5
--
kHz
Remote control and data acquisition
Unit must be user controllable with all measurement parameters available through remote acquisition
--
--
--
--

Line Item 3: Base-Station Emulator

Parameter
Conditions
Min
Typ
Max
Units
Frequency range of operation
Frequency range may exceed typical
--
0.7-2.7
--
GHz

Frequency resolution

--
1
1
Hz

Minimum RF output power

--
-140
-130
dBm

Maximum RF output power

-10
0
--
dBm

RF output level resolution

--
0.05
0.1
dB

Maximum RF input power

+30
+35
--
dBm

Minimum RF input power

--
-40
-40
dBm
Number of RF channels
Capable of supporting 4G LTE with 2x2 MIMO
2
2
--
channels
Supported measurement formats
All conform to 3GPP standards: LTE FDD (data throughput), WCDMA (BER, BLER), GSM (BER)
3
--
--
formats
Remote control and data acquisition
Unit must be user controllable with all measurement parameters available through remote acquisition
--
--
--
--

Line Item 4: Rotating Turntable for Large-Form-Factor Devices

Parameter
Conditions
Min
Typ
Max
Units

Rotating turntable platform size

1.0
1.2
2
m

Rotating turntable platform load

100
200
--
kg
Stepped movement
Angular increment of stepped movement
--
0.1
0.5
deg

Minimum speed for continuous movement

--
0.1
0.5
RPM

Maximum speed for continuous movement

4
5
--
RPM
Control
Remote control required, commands must be made available to user for control with software for other stirring mechanisms such as mechanical paddles and instrumentation such as base-station emulators
--
--
--
--

Calibration, Warranty, Technical Support, Shipping Five years of annual calibration of the channel emulator and base-station emulator is required, preferably through the manufacturer of each instrument.

An extended warranty period for the channel emulator and base-station emulator is desired, but a minimum of 3 years is required.

Basic Technical Support for 1 year

Shipping of all parts