Proposals and/or quotes are NOT being accepted. The solicitation document is linked to Notice of Intent Posting N6264517R0077. The Naval Medical Logistics Command (NMLC) intends to award a Firm Fixed Price (FFP) with General Electric Company to obtain service and maintenance support of digital diagnostic imaging systems.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
RFQ – Service and Maintenance Support of Digital Diagnostic Imaging Systems
This is a combined synopsis/solicitation for commercial items prepared in accordance with FAR 12.6, Streamlined
Procedures for Commercial Items, as supplemented with additional information included in this notice. This announcement constitutes the only solicitation; proposals/quotes are not being requested. The solicitation number is
N6264517R0077 and is linked to NOI Posting N6264517R0077 dated 4 Aug 2017. Provisions and clauses in effect through Federal Acquisition Circular 2005-95 are incorporated. It is the contractor’s responsibility to be familiar with the applicable clauses and provisions. The clauses may be accessed in full text at these addresses:
http://acquisition.gov/far/ and http://farsite.hill.af.mil. The NAICS code is 811219 with a Small Business Size
Standard of $20,500,000.00.
The Naval Medical Logistics Command (NMLC) seeks to execute a Firm Fixed Price (FFP) contract on a sole source basis with General Electric Company to obtain service and maintenance support of digital diagnostic imaging systems located at the Expeditionary Medical Facility (EMF) at Camp Lemonnier, Djibouti, Africa.
ITEM NO
SUPPLIES/SERVICES QUANTITY UNIT UNIT PRICE AMOUNT
0001 12 Months Base Year
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Definium 5000
ECN: 000255; SN: GEHC00374
POP: 30 Sep 17 - 29 Sep 18; Monthly Coverage
FOB: Destination
MILSTRIP: N6264517RC80009
PURCHASE REQUEST NUMBER: N6264517RC80009
ADDITIONAL MARKINGS: EMF DJIBOUTI; N68681-7065-8009; N68681-
17I013
NET AMT
ITEM NO SUPPLIES/SERVICES QUANTITY UNIT UNIT PRICE AMOUNT
0002 12 Months Base Year
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
OEC 9900 Elite
ECN: 000554; SN: E9-2424
POP: 30 Sep 17 - 29 Sep 18; Monthly Coverage
FOB: Destination
MILSTRIP: N6264517RC80009
PURCHASE REQUEST NUMBER: N6264517RC80009
http://acquisition.gov/far/ http://farsite.hill.af.mil/
0003 6 Months OPTION Base/Option
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Optima XR220 AMX
ECN: 000672; SN: SB2201600100WK
Currently under warranty
Start Date: 01 April 2018
POP: 1 Apr 17 - 29 Sep 18; Monthly Coverage
FOB: Destination
0004 6 Months OPTION Base/Option
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Flashpad for Optima XR220 AMX
Subcomponent of Optima XR220 AMX
Currently under warranty
Start Date: 01 April 2018
POP: 1 Apr 17 - 29 Sep 18; Monthly Coverage
FOB: Destination
1001 12 Months OPTION Option Year 1
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Definium 5000
ECN: 000255; SN: GEHC00374
POP: 30 Sep 18 - 29 Sep 19; Monthly Coverage
FOB: Destination
1002 12 Months OPTION Option Year 1
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
OEC 9900 Elite
ECN: 000554; SN: E9-2424
1003 12 Months OPTION Option Year 1
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Optima XR220 AMX
ECN: 000672; SN: SB2201600100WK
1004 12 Months OPTION Option Year 1
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Flashpad for Optima XR220 AMX
Subcomponent of Optima XR220 AMX
2001 12 Months OPTION Option Year 2
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Definium 5000
ECN: 000255; SN: GEHC00374
POP: 30 Sep 19 - 29 Sep 20; Monthly Coverage
2002 12 Months OPTION Option Year 2
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
OEC 9900 Elite
ECN: 000554; SN: E9-2424
2003 12 Months OPTION Option Year 2
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Optima XR220 AMX
ECN: 000672; SN: SB2201600100WK
2004 12 Months OPTION Option Year 2
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Flashpad for Optima XR220 AMX
Subcomponent of Optima XR220 AMX
3001 12 Months OPTION Option Year 3
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Definium 5000
ECN: 000255; SN: GEHC00374
POP: 30 Sep 20 - 29 Sep 21; Monthly Coverage
3002 12 Months OPTION Option Year 3
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
OEC 9900 Elite
ECN: 000554; SN: E9-2424
3003 12 Months OPTION Option Year 3
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Optima XR220 AMX
ECN: 000672; SN: SB2201600100WK
3004 12 Months OPTION Option Year 3
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Flashpad for Optima XR220 AMX
Subcomponent of Optima XR220 AMX
4001 12 Months OPTION Option Year4
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Definium 5000
ECN: 000255; SN: GEHC00374
POP: 30 Sep 21 - 29 Sep 22; Monthly Coverage
4002 12 Months OPTION Option Year 4
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
OEC 9900 Elite
ECN: 000554; SN: E9-2424
4003 12 Months OPTION Option Year 4
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Optima XR220 AMX
ECN: 000672; SN: SB2201600100WK
4004 12 Months OPTION Option Year 4
FFP
Service and Maintenance for Digital Diagnostic Imaging systems
Flashpad for Optima XR220 AMX
Subcomponent of Optima XR220 AMX
PERFORMANCE WORK STATEMENT
1.0. GENERAL INFORMATION
1.1. General. The Expeditionary Medical Facility (EMF) at Camp Lemonnier, Djibouti has requirements for a non-personal services contract to maintain designated Government-owned digital diagnostic imaging systems, manufactured by the General Electric Company and/or their subsidiary companies. Technical Exhibit 1 references the full equipment list.
1.2. Scope. The Contractor shall maintain all equipment listed in the performance work statement. The Contractor shall maintain the systems within the original equipment manufacturers (OEM) specifications, in accordance with all U.S., State, and Local law and regulations, U.S. Navy regulations, instructions, and Joint Commission requirements. The Joint Commission’s requirements are specified in its Comprehensive Accreditation Manual for Hospitals (CAMH), readily found on jointcommission.org. U.S. Navy regulations and instructions will be made available upon request. The contractor must maintain the system such that it meets Navy and DHA
Cybersecurity requirements.
1.2.1. The Firm Fixed Price (FFP) Maintenance shall include unlimited telephone support (24x7), preventive maintenance inspections at a quantity of four (4) per unit per CLIN period of performance, four (4)
Biomedical Equipment Technician service training sessions per system, and two (2) on-site corrective maintenance visits per unit per CLIN period of performance, for all medical devices listed in the performance work statement. All maintenance provisions shall apply to hardware, firmware, and software, as appropriate, unless otherwise stated. The Contractor shall provide all parts, glassware, labor, materials, travel, personnel, supervision, and freight for the scheduled, preventive and corrective maintenance of all equipment identified.
1.2.2. The aggregate amount of radiographic equipment covered under the contract shall fluctuate over time, as hardware and software eclipse useful life and as new equipment comes off warranty. However, the aggregate requirement and level of effort will remain relatively constant over the life of the contract. Any fluctuations to the equipment list will be executed through an official contract modification.
1.2.3. When product recalls or upgrades are necessary to maintain the equipment in service within the original manufacturer’s specifications, the Contractor shall provide all associated hardware and software updates/upgrades.
1.2.4. This contract shall provide unscheduled support/service coverage between the hours of 7:30AM and
5:00PM, local time in Djibouti, Africa, from Monday through Friday, excluding United States Federal
Government holidays. Maintenance and support shall be provided for a one (1) year period of performance spanning 30 September 2017 through 29 September 2018, including four subsequent option years. Vendor shall provide support immediately upon award of a resulting contract.
1.3. Government Points of Contact.
Contracting Officer’s Representative:
To be determined upon contract award.
Technical Liaison:
To be determined upon contract award.
2. DEFINITIONS. As used throughout this work statement, the following terms shall have the meaning set forth below:
2.1. Calibration: This activity should be the procedures for the mechanical adjustments of the device to ensure consistent, standardized performance or output as per the original equipment manufacturer recommendations.
2.2. Contracting Officer (KO): A warranted representative of the federal Government who has the express authority to enter into, administer, and/or terminate contracts and make related determinations and findings.
The KO is the only person authorized to approve changes to any terms, conditions, or requirements of the contract.
2.3. Contracting Officer’s Representative (COR): A Government employee appointed by the KO and responsible for the inspection, evaluation, and acceptance of products and services provided in accordance with the terms and conditions of the contract.
2.4. Contractor Performance Assessment Reporting System (CPARS): Web enabled application that collects and manages the library of automated CPAR’s.
2.5. Corrective Maintenance: Repair, adjustment, calibration, software manipulation and upgrades or other action taken to return a piece of equipment to original manufacturers specifications following a failure, damage, or other incident that causes the product to no longer perform within the original specifications.
2.6. Defense Medical Logistics Standard Support (DMLSS): The automated procurement, maintenance, and supply system utilized to document various material management functions within the command, and throughout the Department of Defense.
2.7. Field Service Report (FSR): Documentation completed by vendor personnel to officially record all instances of contracted troubleshooting, work, and services conducted on equipment. Specifics on required reporting details can be seen in Section 3.9.2.
2.8. Inspections: The activity of doing functional testing and safety inspections of a medical device at regular intervals to ensure that the equipment is safe for use in accordance with manufacturer’s recommended procedures.
2.9. ISO 9001: The ISO 9000 series is a set of five individual, but related, international standards on quality management and assurance. These standards were developed to effectively document the quality system elements to be implemented in order to maintain an efficient quality system in a company. ISO 9001 is the most comprehensive standard covering design, manufacturing, installation, and servicing systems.
2.10. Medical Treatment Facility (MTF) or Patient Care Facility: All gross square footage for building with a primary function of delivering Government health care services. (i.e. hospital or clinic, including all activities providing outpatient and/or inpatient health services for authorized patient care)
2.11. Non-Patient Care Facility: All gross square footage for buildings with a primary focus of non-patient care activities (i.e. storage, warehouse, etc).
2.12. Original Equipment Manufacturer (OEM): The company which has created a specific model of product/device.
2.13. Preventive Maintenance: Preventive maintenance is used to describe those periodic and “as required” services necessary to maintain equipment in an operating condition which minimizes corrective maintenance, including but not limited to calibration, adjustment, lubrication, updating software for security, safety or system functionality issues, as well as that testing required to maintain a safe environment of care which includes electrical safety testing, output testing, replacement of batteries and other shelf life components on a scheduled basis (which becomes corrective maintenance if the component fails due to lack of preventive maintenance).
2.14. Quality Assurance (QA): Actions, including inspections, performed by the Government to determine whether products and services conform to contract requirements.
2.15. Quality Assurance Evaluator (QAE): A Government employee responsible for performing quality assurance functions for a contracted service.
2.16. Quality Control (QC): Actions, including inspections and preparation of inspection records, performed by the Contractor to ensure that products and services conform to contract requirements.
2.17. Scheduled Parts Replacement (SPR): This is the activity of listing of parts that need replacing at certain intervals. This may be either hour-based or date-based, in accordance with manufacturer’s specifications.
3. PERFORMANCE REQUIREMENTS: The Contractor shall provide equipment maintenance services to include all periodic planned maintenance and unscheduled repairs, in accordance with OEM specifications and in compliance with ISO 9001 standards, to include repairs necessitated by operator error. Upon completion of each preventive maintenance visit and each corrective maintenance visit, field service personnel must submit a signed service report, in English, to the COR and Biomedical Technical Liaison for data entry into DMLSS.
The service report shall include all information outlined in Section 3.9.2 of the contract.
3.1. Service Vendor’s Qualification: The Contractor must be an Original Equipment Manufacturer (OEM) certified service provider. Contractors must verify that an individual whom they plan to employ, or continue to employ, in the United States is authorized to accept employment in the United States. Individuals, such as those who have been admitted as permanent residents, granted asylum or refugee status, or admitted in work-related nonimmigrant classifications, may have employment authorization as a direct result of their immigration status.
3.1.1. Field Service Personnel shall be authorized by the OEM-certified service vendor to perform the required maintenance services. The Contractor shall ensure that all field engineers possess current OEM-certification level training on all devices encompassed in the equipment list. Service personnel shall have sufficient experience to demonstrate proficiency in maintenance of the product.
3.1.1.1. Experience of at least three (3) years working on a class of equipment, or over one (1) year on the specific make and model is sufficient to indicate experience. Substantially more time may be required based upon the technology involved. Occasional experience with a limited set of equipment does not meet the requirements of this contract.
3.1.1.2. Service vendors shall provide, to the COR, written assurance of the competency of their personnel as well as a list of credentials of approved field service personnel. The KO and/or COR may authenticate the training requirements and request training certificates or credentials from the service vendors at any time.
3.1.1.3. Government Right of Refusal: EMF Camp Lemonnier reserves the right to reject any of the service vendor’s personnel and refuse them permission to enter the premises or work on EMF Camp Lemonnier equipment. This right may be exercised at any time. The reasons for rejections may include: lack of training and experience, but may also include, but not be limited to: behavior, attire, security concerns, and immigration status.
3.1.1.4. Field engineers shall provide all parts, labor, materials, travel, personnel, supervision, and freight necessary to complete repairs and preventive maintenance on all equipment listed within this contract and shall not invoice EMF Camp Lemonnier for work performed at the direction of the Contractor.
3.1.1.5. Field engineers shall provide services in accordance with the performance requirements specified in this performance work statement.
3.1.1.6. Contractor Field Service Personnel shall hold a current, unexpired, and official United States passport.
3.1.2. Technical documentation: The Contractor shall obtain, have on file, and make available to its field service personnel all operational and technical documentation, which is required to meet the performance requirements of this contract. This documentation shall include any software needed or beneficial in speeding the diagnosis, repair, or calibration of the equipment, equivalent to that used by the manufacturer to perform the same work. The Contractor shall ensure that all Field Service Personnel possess the ability to perform corrective software upgrades and manufacture recall notices, and are in possession of legally licensed diagnostic software necessary for rapid, efficient, and effective maintenance.
3.1.3. The Contractor Point of Contact (POC): The contractor shall provide, in writing, the name and telephone number of a primary and alternate, English-speaking, individual to act as their representative for the scheduling and coordination of service calls.
3.2. Preventive Maintenance: Contractor must insure that 95% or more of all periodic maintenance is performed within thirty (30) days of due date, and that 100% of periodic maintenance is performed within sixty (60) days of the due date.
3.2.1. Equipment in Use: If, for reasons beyond the Contractor’s control, equipment is not made available for maintenance for the entire thirty (30) days before the maintenance due date and sixty (60) days following the due date, the Contractor shall immediately notify the COR.
3.2.2. Equipment Unable to Locate (UTL): If the Contractor is unable to locate an item within the maintenance window specified, the Contractor shall immediately notify the COR. Contractor shall arrange the performance via coordination with the Contracting Officer’s Representative (COR) or Alternate COR during circumstances of unexpected delays (i.e. non-availability of the patient care areas due to room occupancy, etc.), at no additional cost to the Government. The COR may determine that the property has been disposed of or is otherwise inactive or unavailable for use and adjust Technical Exhibit 1, accordingly.
3.2.3. Preventive Maintenance Scheduling: The Contractor shall schedule service vendor preventive maintenance visits as specified for each equipment item. The frequency of preventative maintenance will be four times per year, per system. Preventive maintenance visits shall include complete inspection and testing of the equipment in accordance with DMLSS maintenance procedures identified in DMLSS and/or manufacturer’s recommended guidelines. Preventive maintenance visits shall be scheduled with the EMF
Camp Lemonnier Biomedical Engineering Department at least one month in advance. If the routine scheduling for the equipment items is in excess of 1 month in advance, then the preventive maintenance visits must be coordinated at least that far in advance, so as to not require rescheduling patients.
3.2.4. 100% Preventive Maintenance: Under the circumstances of 3.2.1 or 3.2.2, the COR and Contractor shall arrange a means of complying with the requirements for preventive maintenance at no additional cost to the
Government. Equipment determined missing, lost, or stolen shall be exempt from maintenance.
3.3. Corrective maintenance: The Contractor shall provide for any corrective maintenance services necessary to maintain equipment within the original manufacturer’s specifications, and in accordance with any applicable
FDA requirements. Corrective maintenance shall include damage to equipment that results from, but not limited to, unintentional damage, operator error, normal wear and tear, component part failures, environmental issues (e.g., water damage), etc.
3.3.1. Modifications: The Contractor shall maintain equipment within the original manufacturer’s specifications, and shall make no modifications to the original design.
3.3.2. All “repair by exchange” shall be approved in advance.
3.3.3. Response time: Within three (3) hours of the notification, Contractor technical personnel shall respond by telephone. If the equipment cannot be restored to the operating condition via telephone, the Contractor personnel shall arrive onsite within five (5) business days to complete the requested service. The onsite response time is measured from the time the Contractor is originally called, not from the time the
Contractor returns the call.
3.3.4. Parts: The Contractor shall provide coverage for all repair and replacement parts necessary to maintain the equipment, listed in Technical Exhibit 1, within the parameters of the original manufacturer’s specifications. The Contractor shall supply new, non-refurbished OEM parts for all repairs. Contractor installed replacement parts shall become the property of the Government. The contractor shall extend to the Government all commercial warranties on replacement parts, consistent with standard industry practices. Replaced items shall be properly disposed of at no additional cost to the Government (contractor shall pay all “environmental fees”, where required). All glassware replacement (i.e. x-ray tube, image intensifier) is encompassed in this firm, fixed-price agreement. Replacement parts shall be furnished within five (5) business days. Freight, postage, and storage charges associated with shipment and receipt of replacement parts, and the return of parts, shall be the responsibility of the contractor.
3.3.5. Software maintenance is included in the scope of the contract, including updates and patches as needed.
3.4. No overtime services shall be completed without the prior approval of the COR and/or BIOMED Technical
Liaison. Accordingly, for work determined to be outside of the scope of this agreement, no overtime shall be billed to the Government without prior approval of associated services by the COR and/or KO.
3.5. Recalls and Updates: The Contractor shall provide installation for all manufacturer recalls and mandatory updates within thirty (30) calendar days of the original equipment manufacturer’s notification for all equipment covered under this contract.
3.6. BMET Training: The Contractor shall provide one initial training session for the all MTF Biomedical Repair personnel on the procedures and processes of the Contractor’s maintenance program, immediately upon contract award. This training will provide the Biomedical Technicians insight into the OEM specified procedures for preventative maintenance and frequently observed issues that require corrective maintenance.
3.6.1. For each imaging system covered in this agreement, the vendor shall provide four (4) on-site Biomedical
Equipment Technician service training sessions to the EMF Djibouti Biomedical Equipment Technicians
(BMETs). This service course shall be attended by up to two (2) Department of Defense BMETs. The training sessions can occur during the same on-site visit as those required for preventive maintenance.
Service training sessions will include, but not be limited to: common service error codes, recommended troubleshooting measures, and common user errors and fixes. Due to EMF Camp Lemonnier’s remote location and length of vendor travel time, this training shall be provided to prevent frequent vendor field engineer travel on-site to conduct corrective repairs. The DoD BMET will accordingly be viewed as a service partner of the selected vendor’s field service personnel.
3.7. 100 % Completion During Hours of Work: Work required in the performance of this contract shall be performed Monday through Friday, 0800 – 1700 (East Africa Time Zone), excluding United States Federal
Government holidays. The Contractor shall also provide a telephone consulting service, twenty-four (24) hours per day, seven (7) days a week. This telephone consulting service shall be provided to avoid undue delays in contract service coverage during after-hours and weekend/holiday periods, as applicable. Under no circumstances shall the Contractor assume an inquiry, via telephone or any other means, is authorization to provide services not specifically covered by the terms of the contract.
3.8. Recalls and Updates: The contractor shall provide installation and mandatory updates within five (5) business days for Class I equipment recalls and thirty (3) days for all other classes, upon original manufacturer’s and/or
FDA recall or alert notification. The contractor shall provide documentation of such installation and mandatory updates.
3.9. Service Events: Upon arrival on-site, prior to performing any work, field service personnel shall physically sign in with BIOMED during normal working hours. The COR, Biomedical Technical Liaison, and/or any
Biomed Personnel may accompany the field service person as an observer during any work performed on-site.
3.9.1. Lock Out/Tag Out: The contractor shall personally notify BIOMED, via FSR, of problems that result in the equipment being left disabled upon their departure. If equipment is left disabled, a method of physical use prevention of the disabled equipment (i.e., Lockout, removal to BIOMED) shall be implemented in compliance with Occupational Safety and Health Administration (OSHA) and Navy regulations, specifically Title 29 1910.147 of the Code of Federal Regulations and the OPNAVINST 5100.23G Chapter
24, to prevent any patient, staff, or visitor hazards. Where lockout procedures are not possible, tag-out procedures shall be followed in accordance with the above regulations. Additionally, a label shall be adhered to the disabled equipment specifying the equipment is not serviceable. Reference (i) is available at: http://doni.documentservices.dla.mil/allinstructions.aspx
3.9.2. 95% Service Event Documentation: Upon completion of each preventive maintenance visit and each corrective maintenance visit, field service personnel must submit a signed Field Service Report (FSR) to the COR and Biomedical Technical Liaison. The service report shall indicate all problems addressed, devices checked, parts replaced/repaired, parts ordered, and problem resolution or pending action. Each service report must at a minimum document the following data legibly, in English, and in complete detail:
1. Name of contractor
2. Name of contractor personnel performing service
3. Contractor personnel log/control number
4. Equipment identification factors to include: Manufacturer, make, model, serial number and
Facility Equipment Control Number (ECN)
5. Description of problem reported by user
6. Date & time of initial trouble call
7. Date & time, beginning and ending, of hours on-site for service call
8. Itemized description of service performed to include, but not be limited to: parts used, problem found, corrective action(s) taken or recommended, communications had with
Medical Treatment Facility (MTF) personnel
9. Resolution and/or pending action(s)
10. Estimated total billing cost, if the MTF did not have an existing maintenance agreement:
http://doni.documentservices.dla.mil/allinstructions.aspx
a. Include the list price of services, labor, travel, and parts
b. Note - This pricing information is not to be confused with an invoice
11. Signatures:
a. Contractor performing service
b. DoD BMET verifying service rendered
3.10. 100% Monthly Reports: The Contractor shall submit monthly management reports on the maintenance and repair activity of all equipment under this contract to the COR via electronic e-mail. This report shall address total preventive and corrective maintenance work orders generated, numbers completed as a percentage, and numbers outstanding at the end of each month. Report of previous month activity should be submitted no later than the fifteenth (15 th
) day of the following month.
3.11. Test equipment: Test equipment calibration shall be traceable to National Institutes of Standards and
Technology (NIST). Any efforts conducted without equipment within the calibration intervals is not performance in accordance with this contract. Upon request by COR, the Contractor and/or service vendors shall provide a copy of the current calibration certification of all test equipment to be used, or which has been used, on MTF equipment.
3.12. Removal of equipment from site: No equipment shall be removed without the approval of the COR and
Materials Management Department of the MTF. Property passes for equipment removal shall be obtained by the contractor via the Materials Management Department of the MTF. Any failure to obtain such passes which results in detainment or prosecution of service contractor personnel are the sole responsibility of the service contractor personnel.
3.12.1. Removal Report: The contractor shall provide a detailed description of removed items in writing on the contractor’s company letterhead. For associated items/accessories, contractor shall provide a detailed written description and quantities of items to be removed. The description will include, as a minimum, manufacturer’s serial numbers and equipment control number (ECN) of all equipment / items removed. In the event that items/sub-assemblies/accessories are repaired via “Repair by Replacement” a detailed description of replaced items is to be included upon return of repaired components.
3.12.2. Off-site repairs: No additional charge shall be allowed for work performed off-site, or any additional time required. All charges resulting from a contractor determined requirement to transport Government-owned property, covered by this contract, to and from an alternate repair location shall be the responsibility of the contractor. The contractor shall provide insurance coverage for damage to or loss of equipment while in contractor or service contractor’s custody.
3.12.3. Hard Disk and Solid State Drives (HDD & SSD): In regards to patient information’s confidentiality and privacy, all contractors and/or service contractors shall not remove equipment containing patient information from EMF Camp Lemonnier. HDD nor SSD shall ever be removed from the MTF, by the
Contractor under this maintenance agreement.
3.12.4. Return of Government-owned property: Upon completion of any and all regularly scheduled preventative or unscheduled corrective maintenance, the contractor shall return all equipment owned by the
United States Government.
4. SAFETY REQUIREMENTS: The Contractor shall take safety precautions as the COR or his/her designee may determine to be reasonably necessary to protect the lives and health of occupants of buildings in which work is being performed. The MTF shall notify the Contractor of any non-compliance with the foregoing conditions and the action(s) to be taken. Service vendors shall, after receipt of such notice, immediately correct the conditions to which attention has been directed. If a field service engineer refuses to comply promptly, the
COR may issue an order stopping all or any part of the work. Under such circumstances, the KO can hold the service vendor in contract default. Nothing in this section shall be construed so as to limit the Contractor responsibilities to comply with any existing law, nor to switch jurisdictions.
5. IDENTIFICATION, PARKING, SMOKING MTF REGULATIONS: Field service personnel shall wear visible photo identification at all times while on the premises of the MTF. It is the responsibility of the field service personnel to park in the appropriate designated parking areas. Information on parking is available from
MTF Security personnel. Smoking is strictly prohibited inside any buildings of the MTF. Possession of weapons, alcohol, or pharmaceuticals that are not over the counter and not prescribed to the individual/family is prohibited on the MTF premises. Enclosed containers, including tool kits, are subject to search. Violations of
MTF regulations may result in citation answerable in the United States (Federal) District Court. Violation of any policy which results in lost time for Contractor is the sole responsibility of the Contractor, and shall not be charged to the Government.
6. GENERAL LIABILITY INSURANCE: The Contractor shall be required to have liability insurance to cover any actions on the part of the field engineers which result in injury or death to patients, staff, visitors, or other
Contractors. Such comprehensive form of policy shall provide coverage of at least $500,000 per occurrence
(FAR 28.307-2).
7. PROPERTY DAMAGE LIABILITY INSURANCE: Service vendors are required to have Property Damage
Liability coverage of at least $500,000.
8. MODIFICATIONS TO CONTRACT: Only the KO has the authority to obligate the Government, and then only in writing. No modification or instruction from any party shall be construed to authorize the Contractor to take any action for which they expect compensation without written authorization from the KO.
9. QUALITY ASSURANCE SURVEILLANCE PLAN (QASP)
Performance
Objective
Performance
Standard
Acceptable
Quality Level
Monitoring
Method/Performance
Assessment
PWS
Reference
Medical equipment essential to the healthcare delivery mission is properly maintained ensuring a safe environment of care for patients and staff alike.
Preventive maintenance for all medical equipment under this contract is completed on or before the scheduled date.
95% of all periodic, preventive maintenance is performed within 30 days of the due date.
100% of all periodic, preventive maintenance is performed within 60 days of due date.
Contractor service reports, DMLSS data sampling by COR, and random physical spot checks by BMET personnel.
3.2
Accurate and timely submission of information critical for effective program management and contractor performance evaluation.
Management Reports of previous month activity should be submitted NLT 15th of the following month.
95% completion of service reports related to equipment under contract.
To include UTL, PM completion percentage, Awaiting Parts, and
Operator Errors.
COR shall monitor date of receipt and completeness of service reports.
3.10
Response time for corrective maintenance calls.
Telephone contact within 3 hours and on-site within 5 business days.
Contractor shall be compliant 95% of the time for both telephone and on-site responses.
COR shall review log of calls and verify response times via DMLSS documentation.
3.3.3
Equipment critical to healthcare delivery is operational and available for use as required.
Equipment listed in Technical
Exhibit 1 shall be fully functional.
Contractor is expected to return the device to service.
COR shall review DMLSS data for compliance.
1.2.3
Equipment post-service event status Upon completion of each preventive maintenance visit and each corrective maintenance visit, field service personnel must submit a signed service report to the
COR or Biomed.
Contractor shall be compliant 95% of the time for providing signed service report to COR.
COR shall review DMLSS data, service log and service reports for compliance.
3.9.2
All work to be completed during normal work hours.
No work outside of established work hours.
100% completion of work during normal work hours.
COR shall review DMLSS data, service log and service reports for compliance.
3.7
9.1. Compliance with Applicable Laws: Contractor shall be knowledgeable of and comply with all applicable
Federal, State, and local laws, regulations, and requirements regarding the conduct of his business.
9.2. Site Visits. The Contractor or his representative having authority to bind the Contractor, shall be required to visit the site at least semi-annually and meet with the KO and/or COR, as required, to discuss and develop mutual understanding relative to safety precautions, labor provisions, and other contractual requirements. This requirement can be cancelled only at the discretion of the KO and/or COR. The Government reserves the right to schedule meetings to discuss or resolve issues/problems, as it deems to be critical, at no additional cost to the
Government.
9.3. Work Performed by Others. When work performed by the Government or other Contractors is expected to affect the Contractor’s performance of work, the Government shall coordinate with the Contractor and provide necessary information (plans, specifications, etc.) for such work to the Contractor such that the Contractor may make scheduling allowances. The COR shall resolve any schedule conflicts between the Contractor and work performed by the Government or other Contractors. Such actions shall not result in additional cost to the
Government under this contract.
9.4. Government Right to Change Standard Operating Procedures. The Government reserves the right to establish and/or change Standard Operating Procedures which shall be binding on the Contractor, i.e., work order procedures, emergency and fire procedures, security procedures and power failures.
9.5. Phase-In Service: Upon award, which should occur prior to expiration of existing service contracts, the
Contractor shall have the opportunity to inspect the existing equipment and note any deficiencies. This action shall occur in sufficient time to notify the existing contract holder to correct deficiencies noted (at least 24 business hours prior to expiration of contract). Failure to submit such list in a timely manner shall be considered acceptance of the equipment in the existing condition. The Contractor shall not be eligible to claim expenses to bring the equipment into manufacturers’ specifications.
9.6. Phase Out of Service: Upon expiration or termination of this contract, the Contractor, at the Contractors’ expense, shall provide OEM certifications for all equipment covered under contract at the rates/discounts normally provided to Government customers. If repairs or corrections to the equipment have to be made to obtain such certifications, the contract shall arrange to have those repairs or corrections made at the
Contractors’ expense.
10. CONTRACTOR-FURNISHED ITEMS AND SERVICES:
10.1. Government Responsibility for Damage. The Government shall not be responsible, in any way, for damage to the Contractor’s stored supplies, material, equipment, or to Contractor’s employees’ personal belongings brought onto the site.
10.2. Physical Security. The Contractor shall safeguard all Government property provided for Contractor use.
The Contractor shall designate an individual and alternate(s) to receipt and account for Government-furnished facilities and equipment.
10.3. Contractor shall furnish all supplies, materials, repair or replacement parts, tools, equipment software upgrading and labor necessary to efficiently and satisfactorily perform the requirements of this contract.
10.4. The Contractor shall protect all facilities from damage caused by the Contractor and shall repair, at no additional cost to the Government, all damages caused by the Contractor’s activities. During performance of services, Contractor shall keep EMF Camp Lemonnier safe, neat, clean and orderly. Contractor shall be responsible for all means, methods, techniques, sequences, and procedures of the services. Contractor employees shall demonstrate appropriate behavior and dress at all times.
10.5. Contractor Supplies and Equipment. The Contractor supplies and equipment shall comply with Federal, State, municipal, and industry or OEM specifications.
10.6. Two-way Radios & Cell Phones. The use of Contractor provided two-way radios and cell phones for communication between personnel must be approved by the Base Communications activity and may not be used in designated areas in EMF Camp Lemonnier. Contractor shall submit a written request to use the radios to the COR. The request shall detail proposed equipment, operating frequencies, and other particulars in writing.
The Government shall not be liable for any damages as a result of not permitting the use of radios within the facility.
10.7. The Contractors personnel shall comply with the DoD Information Assurance Policies applicable.
10.8. Contractors are required to maintain a Drug Free Workplace.
11. OCCUPATIONAL HEALTH.
11.1. Occupational Health/Immunizations: Contract shall comply with applicable Federal and State occupational disease statutes. The contractor shall be immunized annually with the influenza vaccine with the contractor bearing the total cost. On an annual basis, contractor shall provide a current Purified Protein
Derivative (PPD) reading or an evaluation if they are a known PPD reactor. The contractor is responsible for any expenses incurred for required testing. Contractors are required to maintain a Drug Free Workplace, as well as to require any sub-contractors to do the same.
12. PERSONNEL SECURITY.
12.1. Contractor Requirements for Background Checks: It is the contractor’s responsibility to ensure its employees undergo proper criminal history background checks by providing the Government support, cooperation, and active assistance throughout the process.
12.1.1. The contractor shall ensure that employment applications for potential healthcare workers contain a question asking whether the individual has a criminal history, and if so, requiring a description of the disposition of the arrest or charge. The application shall further state that it is being signed under penalty of perjury, with the applicable Federal punishment for perjury stated on the application.
12.2. Government Requirements for Background Checks: The contractor shall comply with DoDD 8500.1, “Information Assurance (IA)”, DoD Instruction (DODI) 8500.2 “Information Assurance (IA) Implementation”, DoDD 5400.11, “DoD Privacy Program”, DoD 6025.18-R, DoD Health Information Privacy Regulation and
DoD 5200.2-R, “Personnel Security Program Requirements”. Contractor responsibilities for ensuring personnel security include, but are not limited to meeting the following requirements: Follow the Privacy Office guidelines for submittal of IT security clearances and ensure all contractor personnel are designated as IT-I, IT-
II or IT-III where their duties meet the criteria of the position sensitivity designations.
12.2.1. Because of the unique circumstances presented by DoD and DON networks, personnel security requirements shall be followed to ensure appropriate precautions are taken prior to allowing contractor personnel access to the network. Any contractor personnel that will be accessing the medical device/system while installed on the hospital network will be required to have a National Agency Check
(NAC) completed. Typically, this requires an investigation to support a “Public Trust Position” and requires the person(s) to complete and submit a Standard Form 85P (SF85P), Questionnaire for Public
Trust Positions, via the Electronic Personnel Security Questionnaire (EPSQ). Questions relating to SF85Ps and the EPSQ process may be directed to 1-888-282-7682 or online at http://www.dss.mil/index.htm.
Contractor personnel accessing equipment connected to the hospital network will be required to complete a
System Authorization Access Request-Navy (SAAR-N) (form OPNAV 5239/14). Copies of this form can be obtained from the Naval Medical Logistics Command (NMLC), or the local MTF. Additionally, contractor personnel are required to complete the annual DoD IA training requirements.
12.2.2. The contractor shall initiate, maintain and document personnel security investigations appropriate to the individual’s responsibilities and required access to Sensitive Information (SI).
12.2.3. Immediately report to the appropriate Navy POC and deny access to any automated information system
(AIS), network, or information if a contractor employee filling a sensitive position receives an unfavorable adjudication, if information that would result in an unfavorable adjudication becomes available, or if directed to do so by the appropriate Navy representative for security reasons.
12.2.4. Ensure that all contractor personnel receive DoD IA training before being granted access to DoD AIS’s.
13. RISK MANAGEMENT FRAMEWORK (RMF)
13.1. The contractor shall comply with RMF requirements, as specified by the Defense Health Agency (DHA) that meet appropriate DoD, DHA and Navy Cybersecurity requirements. The contractor shall initiate the process by providing the required documentation necessary to submit their system to the Authorizing Official
(AO) for assessment and authorization (A&A) for all orders delivered against this contract. The requirements shall be met before the contractor's system is authorized to access DoD data or to interconnect with any DoD network or system that receives, processes, stores, displays or transmits DoD data. The contractor shall ensure the proper contractor support staff is available to participate in all phases of the RMF process. This includes, but is not limited to;
13.1.1. Attending and supporting RMF and A&A meetings with Navy and DHA cybersecurity representatives as required.
13.1.2. Support/conduct vulnerability mitigation to comply with cybersecurity controls listed in National Institute of Standards and Technology (NIST) Special Publication (SP) 800-53; Security and Privacy Controls for
Federal Information Systems and Organizations and DHA Administrative Instruction 77. In order for a system to be authorized, there shall be zero Category 1 findings.
13.1.3. Support the A&A Team during system security testing.
13.1.4. Once the system has been authorized, the contractor shall maintain an acceptable security baseline consistent with the authorization granted by the AO throughout the life of the system. This includes ensuring that all vulnerabilities affecting the operating system, primary application, database and all third party software have validated and applied within 30 days.
13.2. Cybersecurity: The vendor shall establish appropriate administrative and technical safeguards to protect any and all data, to ensure the confidentiality, integrity, and availability of Government data under their control.
At a minimum, this shall include provisions for vendor personnel security and electronic security.
13.2.1. The proposed system shall be capable of supporting the use of Defense Information Systems Agency
(DISA) approved antimalware applications. The Navy uses McAfee Host Based Security System (HBSS) as their solution. The vendor shall provide technical specifications that clearly demonstrate whether the proposed solution can integrate and support, either fully or partially the operation without performance degradation of the medical system/device. In cases where the operation of HBSS is not technically achievable, the vendor shall provide detailed justification and a Plan of Actions and Milestones (POA&M) describing steps towards compliance with this requirement.
13.2.2. The vendor must provide their Antivirus (AV) policy and a list of AV Exceptions (if applicable) to the
NMLC prior to AV installation on the system/device. AV must be installed on all devices before they are placed on the Navy network.
13.2.3. Systems shall be configured in such a way to allow for the updating of malware definition signatures on a scheduled basis. Scanning shall encompass the entire system (file system, operating system, real-time processes), by default. In cases where scanning of the entire system may negatively affect its operation, the vendor shall provide a detailed list of exclusions with justifications.
13.3. Vulnerability Management: Vulnerability mitigation strategies include security updates, service packs, and changes to operating procedures as physical and cyber vulnerabilities are detected. Operating systems, servers, development platforms and the application being delivered to the Navy shall be in compliance with all known applicable guidelines. The vendor shall agree to be proactive in working with the Navy to mitigate new threats as they emerge.
13.3.1. All vendor validated and approved patches, up to and including the date of install, must be loaded on the system prior to system go-live. Security scans will be run during Government acceptance testing. All
Category I (CAT I) findings must be remediated prior to Government acceptance of the system. A CAT I vulnerability is any vulnerability, the exploitation of which will, directly and immediately result in loss of
Confidentiality, Availability, or Integrity. There shall be no more than 10 Category II (CAT II) findings at the time of Government acceptance. CAT II vulnerabilities are any vulnerability, the exploitation of which has a potential to result in loss of Confidentiality, Availability, or Integrity.
13.4. Domain Name System Realm/Directory Services: Contractor will be required to provide technical evidence, if applicable, whether client/server topology based medical systems can integrate with Directory
Services and support LDAP authentication.
13.5. Local Privileged and Administrative User/Local System Accounts: Contractor shall create a single local user account with administrative/root level privileges for purposes of conducting system repairs and maintenance only. This account shall be separate and distinct from the built-in local administrative/root account provided by the Operating System and shall comply with DoD policy. All factors required to complete successful identification, authentication and authorization against the built-in local Administrative/Root level account shall be provided to the MTF Biomedical Engineering Department.
13.5.1. Complete administrative system rights shall be provided to the Government System Administrator for the purpose of conducting device vulnerability scans as needed.
13.6. DHA Business to Business (B2B) Gateway: All contractor systems that will communicate with DON systems will interconnect through the established DHA Business to Business (B2B) gateway.
13.6.1. Contractors will connect to the B2B gateway via a contractor procured Internet Service Provider (ISP) connection and assume all responsibilities for establishing and maintaining their connectivity to the B2B gateway. This will include acquiring and maintaining the circuit to the B2B gateway and acquiring a
Virtual Private Network (VPN) device compatible with the DHA VPN device. Maintenance and repair of contractor procured VPN equipment shall be the responsibility of the contractor.
13.6.2. Contractors shall configure their network to support access to Government systems (e.g., configure ports and protocols for access).
13.6.3. Contractors shall provide full time connections to a TIER1 or TIER2 ISP. Dial-up ISP connections are not acceptable.
13.6.4. Contractors will comply with DoD…