1. Question - How long do they want the High Voltage Cables for this unit.
Answer - The HV cable length needs to be the distance from the tube head to the anode and cathode generators. The generators appeared to be under the cabinet.
2. Question - "This price includes completed cabinet painted with epoxy paint, an adjustable FFD up to 60 inches" What is the range of FFD from 60inches down to?
Answer - The distance from the focal spot to the film should be able to go as high as 60 inches. If the film is laid flat on the interior deck of the cabinet, the focal spot should be able to be 60 inches higher that the film.
3. Question - What is the clear opening size of the hinged door?
Answer - door opening of 24 inches wide minimum.
4. Question – solicitation referenced a picture but no picture was provided.
Answer – picture now attached.