Statement of Work APN7-17 Requirement

A. REQUIREMENTS:

1. Mfr. one year warranty with sticker affixed to each unit plus a warranty certificate.

2. Calibration Certificate and calibration sticker (as applicable for each unit)

3. Units shall be serialized.

4. Units (over $5,000) shall have an IUID Label applied per DFAR 252.211-7003.

5. Manufacturer’s Standard Operating and servicing manuals (CD version acceptable) shall be provided with each unit. The DFARS clause listed below must be referenced in all contracts and the manufacturer made aware that the use/reproduction of tech data/manuals (for government use only) will follow these use guidelines.

DFARS SUBPART 227.71 - Rights in Technical Data 7.

The US government shall have limited technical data rights for all technical manuals and documents furnished with the procured commercial items, for the purpose of Government use only in duplicating or posting technical documentation to limited access web sites for use by US Government Personnel and Contractors engaged in maintenance, calibration or related logistics activities associated with the procured commercial items in accordance with DFARS 252.227-7015 Technical Data--Commercial Items

B. Tasking:

The contractor shall provide the following instrument upgrade:

1) P/N 5730A High Performance Multifunction Calibrator Qty 2 To include the following:

1. Temperature Performance:

a. Operating: 0°C to 50°C

b. Calibration: 15°C to 35°C

c. Storage: -40°C to 75°C

2. Relative Humidity:

a. Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C

b. Storage: <95 %, non-condensing.

3. Safety: IEC 61010-1: Overvoltage Category II, Pollution Degree 2

4. Operating Altitude: 2000 m maximum

5. Guard Isolation: 20 V

6. Electromagnetic Environment: IEC 61326-1: Controlled

7. Line Power:

a. Line Voltage: 100 V-120 V, 220 V- 240 V ▒10 %

b. Line Frequency: 47 Hz-63 Hz

c. Maximum Power: 300 VA

8. Weight: 27 kg (62 lb.)

2) P/N 5730CASE Transit Case Qty 2 To include the following:

1. All materials should be RoHS compliant to directive 2002-95-EC and revisions.

2. Supplier will provide certificate of compliance and constituent materials.

3. No sharp edges

4. Foam: 1780 polyurethane 4# x PE

3) P/N 8508A Calibration Kit Qty 20 To include the following:

1. 1 GΩ resistor: 100 ppm/yr. and 25 ppm/°C stability, factory calibration uncertainty 100 ppm (95% CL) Qty 1 each kit

2. Screened lead set that connects 8508A reference Multimeter to calibrators Qty 1 each kit

3. Four wire shorting PCB (used to null front and rear input) Qty 2 each kit

4. Carrying case Qty 1 each kit

4) P/N RPM4-LP Low Pressure Monitor Qty 5

To include the following:

1. Power requirements: RPM4: 85 V ac to 264 V ac, 50/60 Hz, 25 V a max and 12 V dc @ 9 Ah Battery/charger: 100 V ac to 240 V ac, 50/60 Hz

2. Normal operating temperature range: 15 °C to 35 °C (59 °F to 95 °F)

3. Vibration: Must meet MIL-T-28800D

4. Weight: 5 kg (11 lb.)

5. Dimensions (H x W x D): RPM4: 10 cm x 22.7 cm x 24 cm (3.9 in x 8.9 in x 9.5 in)

Battery/charger: 8 cm x 22.5 cm x 20 cm (3.1 in x 8.9 in x 7.9 in)

6. Communications ports: RS-232 (COM1, COM2), IEEE-488.2

7. Maximum/minimum working pressure: ± 15.6 kPa (± 62.4 inH2O)

8. Pressure range: -15 to 15 kPa (-60 to 60 inH2O)

9. Operating media: Clean, dry, noncorrosive gas

10. Calibration: ISO 17025 Accredited (A2LA) Calibration Report to be included

11. Pressure connections: 1/8 in. NPT F

12. Resolution: To 1 ppm, user adjustable

13. Predicted one year stability: ± 0.005 % of reading

14. Measurement uncertainty for 1 year: ± 0.015 % of Auto Ranged span

15. Minimum auto ranged span: ± 1.25 kPa (± 5 inH2O)

5) IUID LABEL Qty 7 C. Deliverables Delivery of the equipment shall be provided at the following address:

COMMANDER

NAVAIR Calibration Standards

MCAS, Attn: Lou Cerrillo

(UIC N48535)BLDG 612, Bay 9

Beaufort, SC 29904-5002 D. Delivery Date Delivery shall be no later than 6 weeks after award of contract, and shall be absolutely no later than November 01, 2017.

***PARTICAL SHIPMENT ACCECPTED***

E. Invoicing All invoicing shall be submitting into Wide Area Workflow for payment after ALL material is received in the condition as stated in section B.