Statement of Work for

High Resolution Computed Radiography System 1.0

SCOPE

The Contractor shall provide a computed radiography (CR) system for use in ordnance inspection at the Naval Surface Warfare Center, Indian Head Explosive Ordnance Technology Division (NSWC IHEODTD). The CR system will be used with energy levels of 60 kV up to 320kV. The CR system will supplement an existing CR scanner so must be the same model since system is already qualified. The CR system shall include an acquisition computer system that can be installed in the existing control console.

2.0

APPLICABLE DOCUMENTS

ASTM E2339 - Standard Practice for Digital Imaging and Communications in Nondestructive Evaluation (DICONDE) ASTM E2445 - Standard Practice for Qualification and Long-Term Stability of Computed Radiography Systems ASTM E2698 – Standard Practice for Radiological Examination using Digital Detector Arrays

3.0

SPECIFIC SYSTEM PERFORMANCE REQUIREMENTS

The Contractor shall provide a CR system with all hardware and software necessary to capture, display, analyze, store and send digital x-ray images for industrial radiographic inspection. In addition to meeting these requirements, the system shall meet the specific requirements of Statement of Work (SOW) Paragraphs 3.1 to 3.5.

3.1 Computed Radiography Scanner shall meet the following criteria:

· Manufacturer: DURR

· Model: HD-CR 35 NDT

· Imaging Plate Size: up to 14” x 17”

· Scanning Resolution: Variable, 100 micron and at least one setting less than 25 micron

· Laser Spot Size: less than 25 micron

· User adjustable PMT voltage and laser intensity

· Capable of sample binning, including generating a 50 micron image from 25 micron sampled data and a 25 micron image from data sampled at less than 25 micron.

· Bit Depth: 16-bit, 65,536 gray levels minimum

· Throughput (for 24cm x 18cm, 25 micron resolution): 10 imaging plates/hour minimum

· Throughput (for 14” x 17”, 50 micron resolution): 15 imaging plates/hour minimum

· Touchscreen display and memory card storage to allow operation of scanner without computer system

· Erasure: Built-in and automatic 3.2 Imaging Plates shall meet the following criteria:

· Ultra-Resolution imaging plates, optimized for scanning at 25 microns or less. Twelve (12) 14 x 17 inch imaging plates shall be provided.

3.3 Image Acquisition Station shall meet the following criteria:

· Standalone desktop system for capturing digital data from the CR scanner, storing and viewing images, including all computer hardware and software required.

· Processor, Intel Zeon E5-1650 v4, 3.50 GHz or equivalent

· RAM, 32GB DDR4 minimum

· Operating system shall be Windows 10 Pro

· System shall include a Trusted Platform Module (TPM) version 1.2 or higher.

· System shall not include solid state hard drives

· System shall not contain any wireless, Wifi or 802.11 device.

· System must be DICONDE compliant

· Dual monitor, both must be VESA mountable.

· High resolution Medical Grade 3 Megapixel, 21” flat display with 10-bit display minimum. EIZO Radiforce RX340 or equivalent. Monitor shall be compliant with DICOM Part 14. Monitor shall contain a backlight brightness sensor and automatically adjust brightness. Brightness shall be 250 cd/m2 minimum and contrast shall be 250:1 minimum.

· High Resolution 20” to 22” 1080P (1920x1080) monitor.

· Image display shall meet all requirements of ASTM E2698, section 7.5. System must be capable of displaying a full screen test pattern on the Medical Grade monitor in accordance with SMPTE RP 133 (or equivalent) to allow performance of those tests.

· Graphics card to support full resolution and bit depth of medical grade monitor. EIZO MED-X4900 or equivalent.

· External USB DVD burner 3.4 Acquisition and Viewing Software shall be D-Tect version 7.3.0 or newer and meet the following criteria:

· Digital images shall be stored in the DICONDE format with the following metadata fields as specified by the user at a minimum:

· Component ID (0010,0020)

· Component Name (0010,0010)

· Study Description (0008,1030)

· Study Date (0008, 0020)

· Series Description (0008, 103E)

· Examining Company’s Name (0008, 0090)

· System shall save the following metadata fields automatically:

· Manufacturer (0008, 0070)

· Manufacturer’s Model Number (0008, 1090)

· Device Serial Number (0018, 1000)

· Modality (0008,0060)

· Imager Pixel Spacing (0018, 1164)

· Software Version(s) (0018, 1020)

· Original (raw) image without any image enhancements shall be retained. The original data shall be capable of being re-processed and re-displayed using the same or other process algorithm without having to take another radiographic exposure.

· System shall allow the user to select images for viewing from a list. The list shall show the following fields at a minimum and shall allow the user to sort the list by any of these fields. User shall also be able to filter the list to only show specific values for any of these fields.

· Component ID

· Component Name

· Study Date

· Examining Company’s Name

· Image viewing shall allow for the following:

· Images shall be initially displayed with no image enhancement and then allow the user to apply filters, edge enhancements or contrast enhancements as desired.

· User adjustable zoom. Minimum zoom shall allow the viewing of the entire image on the screen. Maximum zoom shall be a factor of at least 8 times greater than the minimum. User must be able to view the image with one pixel on the monitor equal to one pixel in the scanned image.

· User adjustable window and level. The range of gray values currently being displayed relative to the entire range of gray values present in the image shall be shown graphically. The gray value for window width and the window center shall be displayed. Window width and window center shall be adjustable independently by the user.

· Application of a contrast enhancement algorithm to the displayed image with user adjustable levels or settings.

· Application of filters including smoothing, sharpening, unsharp masking and edge enhancement.

· Single or multiple images shall be viewable at one time. User shall have the ability to change the viewing parameters of each image independently or all images simultaneously.

· User shall have the ability to save multiple views of an image. User shall be able to recall these views at a later date. Stored views shall include window and level, zoom in addition to any enhancements or filters.

· All image manipulations and calculations shall be conducted with at least 14-bit grayscale pixel depth.

· Image viewing interface shall provide buttons that allow the following functions to be quickly and easily performed, preferably with a single mouse click.

· Setting zoom level to display the entire image on the screen.

· Setting zoom level to common zoom levels such as 2X and 4X

· Rotation of images in a clockwise or counterclockwise direction in 90 degree increments

· Flipping images vertically and horizontally.

· Inversion of the gray scale of the displayed image.

· Automatic adjustment of window and level to match the range of gray scale values present in a region selected by the user.

· Image analysis tools shall include the following:

· Gray scale values for all image analysis tools shall be linear.

· Histogram of a user-selected area with statistics including mean, standard deviation, minimum, maximum gray values. Additionally, height and width of a rectangular area (in pixels) and total pixel count shall be displayed.

· Line profile tool with ability to generate an integrated or averaged profile for a line of user-selected width.

· Signal-to-Noise (SNR) of a user-selected area

· Linear measurement tool capable of being calibrated to an object of known size in the image.

· Image mathematical operations including addition and subtraction.

· Point measurement tool to display the gray value of a single pixel or very small area. Tool shall have the ability to move the point across the image with the mouse and display the gray value of this point in real time.

· Images shall be exportable to 16-bit TIFF and JPG formats

· User shall be able to export the entire image or part of an image. Selection of the part of the image to export can be by selecting a region of the image or by exporting the section of the image currently displayed on the screen.

· User shall be able to export the original data in addition to being able to export the image adjusted for window/level and any enhancements applied.

· Images shall be writable to a DVD in the DICONDE format.

· DVD shall include a viewing program that allows viewing of the images on a computer that does not have any special, proprietary software installed. Viewing program shall be capable of displaying the image with and without the image enhancements views saved on the original workstation.

3.5 Supplemental Material The Contractor shall provide Supplemental Material that meets the following criteria:

· Lithium-ion battery for powering scanner

· All cables required shall be provided by the contractor

· Cleaning material for imaging plates, sufficient to clean at least 24 imaging plates

· Any cleaning material and/or cleaning tools required to perform user maintenance on scanner.

· Shipping/storage case for scanner. Cases shall be hard-shell and sufficiently padded to minimize the potential for shipping damage. Case for scanner shall have wheels. Case shall allow scanner to be operated from base of case with only lid removed.

4.0

WARRANTY

The Contractor shall provide a standard comprehensive 1-year warranty.

Enclosure E – Statement of Work