This document is an attachment to a federal contract opportunity, providing detailed requirements for a high-resolution, variable pressure Schottky field emission scanning electron microscope (SEM) system. The key requirements include:
The SEM system must have a variable pressure capability, able to operate in both high and low vacuum conditions. It must have a field-emission gun electron column capable of high spatial resolution (less than 1.3 nm at 1 keV, less than 0.9 nm at 15 keV in high vacuum). The system must be equipped with various electron detectors and an automated 4-axis stage to accommodate large samples up to 8 kg. An automated load-lock system compatible with the NRL's custom sample pucks is required. The SEM must include fully scriptable operation via a Python API, with automated features such as auto focus and auto contrast. The awarded contractor will be responsible for installation, training, and a 3-year service contract.
The related federal contract opportunity is a sole-source solicitation (N00173-24-Q-MR23) from the Department of the Navy's Office of Naval Research for the procurement of this TESCAN SEM system. Responses providing evidence of competitive alternatives are due by September 20, 2024.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
The Materials Science and Technology Division at the US Naval Research Laboratory (NRL) is in need of a high-resolution, variable pressure Shottky field emission scanning electron microscope (SEM) that is capable of automated operation, including a automated sample loading/unloading load-lock system. Further requirements are listed below:
1.) The system must include a variable pressure system that can image both at high vacuum conditions (less than 1x10–3 Pa) or in a low vacuum condition (7 – 500 Pa).
2.) The system must be equipped with a field-free ultra high resolution scanning electron column capable of:
a.) Accelerating voltages from 200 eV (or lower) to 30 keV (or higher).
b.) Spatial resolutions that are less than/equal to 1.3 nm at 1 keV; less than/equal to 0.9 nm at 15 keV in high vacuum mode. With low vacuum mode, the resolution should be less than/equal to 2.0 nm at 30 keV.
c.) Probe current range that is adjustable from less than/equal to 2 pA to greater than/equal to 400 nA.
d.) Capable of achieving a distortion free field of view of greater than/equal to 4.5 mm at a working distance of 10 mm, and greater than 50 mm at maximum working distance.
3.) Must be equipped with detectors and imaging system consisting of:
a.) Secondary electron (Everhart-Thornley) detector.
b.) Motorized retractable annular back-scattered secondary-electron detector (with one or more detection zones).
c.) Imaging system capable of greater than/equal to 16,000 x 16,000 pixel frame sizes with simultaneous channel data collection.
And may be equipped with d.) In column energy filtered detector.
e.) In column axial detector.
4.) The system must include a motorized 4-axis stage capable of:
a.) X,Y travel greater than/equal to 130 mm.
b.) Z axis travel greater than/equal to 100 mm.
c.) Tilt range from less than/equal to –70° to greater than/equal to 90°.
d.) Accommodating a max specimen height of 130 mm or greater.
e.) Accommodating a max specimen weight of 8 kg or greater.
f.) Active vibration isolation system.
5.) The system must have an automated load-lock that includes:
a.) Automatic loading/unloading to/from the main analysis chamber.
b.) The load-lock must be compatible with the NRL Robomet Sample pucks. These are custom designed pucks, 50 mm in diameter, 25 mm in height, that carry samples that are an additional 20–30 mm in height. These pucks integrate with other processing/analysis equipment. Design drawings can be provided on request.
c.) Placement of the NRL Sample pucks onto the SEM stage must have a precision of less than 50 microns in X-Y, with less than 10 micron prefered.
6.) The system must be equipped with automation software that includes:
a.) Fully scriptable operation of the microscope through a Python API that must include, automatic sample loading/unloading through the load-lock, vacuum system control, electron column controls for all scanning modes, detectors settings and positions, brightness/contrast, focus, stage movements, beam voltage, beam current, magnification, and imaging parameters. The API must include full compatibility with the already developed NRL scripting routines based on the SharkSEM
API.
b.) Additional automation features must include at least, auto focus, auto contrast/brightness, auto sigmation correction, objective centering, and sigmator centering.
7.) The system installation shall include:
a.) Environmental survey of the planned installation room.
b.) Delivery of the system, and installation and checkout of SEM operation.
c.) Transfer of the already acquired EDAX Inc Octane energy dispersive spectroscopy (EDS) and electron backscatter diffraction (EBSD) detectors from NRL’s current SEM to this new SEM.
d.) Two day onsite training of SEM operation.
8.) The system must have a service contract that spans three years and includes:
a.) labor, travel, living expenses and replacement parts coverage (including system software and/or software updates applicable to current configuration) b.) One preventive maintenance visits per year c.) Unlimited technical support via email, phone or web.
File details come from the government source that posted it. Updated .