TECHNICAL REQUIREMENTS
FOR
TIME DOMAIN TERAHERTZ SPECTROMETER SYSTEM
The Naval Research Laboratory is seeks to acquire a Time Domain Terahertz (THz) Spectrometer system, to be coupled to an existing all-reflective microscope Hyperion 1000 Bruker Optics.
Specifications that shall be met by the Time Domain Terahertz Spectrometer system are:
1. The spectrometer shall cover from 0.1 to more than 5THz.
2. The spectrometer shall be computer controlled. The software and computer to run the spectrometer shall be included with the system.
3. The spectrometer shall be controlled through Labview. The Labview drivers shall be included with the system.
4. The spectrometer shall have fiber-coupled antennae for easy integration for both emitter and receiver.
5. The average terahertz power shall be 25 µW or higher.
6. The peak dynamic range shall be larger than 90 dB.
7. The spectrometer shall include a pulsed laser for the generation of THz light with a pulse width shorter than 60 femtoseconds.
8. The spectrometer’s resolution shall be better than (meaning <) 5GHz at 200 ps and better than (meaning <) 40 GHz at 20 ps.
9. The spectrometer shall include a computer-controlled mechanical delay stage.
10. The spectrometer shall have a usable terahertz path length of 15 to 80 cm; and
11. Delivery, installation and demonstration shall be included.