20Q0099_sol_0001.pdf
PDF 393 KB Posted
- Attached to
- LIQUID CHROMATOGRAPHY/MASS SPECTROMETER Federal contract opportunity
- Solicitation number
- N0016420Q0099
About this file
This document provides details for a federal government solicitation seeking an Ultra Performance Liquid Chromatography Photodiode Array/Mass Spectrometer. The Naval Surface Warfare Center is requesting quotes for a UPLC/PDA/MS in accordance with the salient characteristics specified, including capabilities such as high mass resolving power, dynamic range, and ability to perform MS/MS product ion scanning. Quotes are due by April 20, 2020 at 12:00 PM Eastern Time and are to be emailed to the point of contact. The solicitation is set aside for small businesses and is being conducted according to FAR procedures for commercial items. The Navy intends to evaluate quotes and potentially negotiate a contract award.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 20Q0099_sol.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER (Type or print)
30-105-04EXCEPTION TO SF 30
APPROVED BY OIRM 11-84
STANDARD FORM 30 (Rev. 10-83) Prescribed by GSA
FAR (48 CFR) 53.243
This amendment is to clarify the follow ing language w ithin Salient Characteristics From: Multiple day training for at least four people must be offered by the manufacturer for both the UPLC/PDA and UPLC/PDA/MS/MS systems.
To: Multiple day training for at least four people must be offered by the manufacturer for both the UPLC/PDA and UPLC/PDA/MS/MS systems.
Training for the UPLC/PDA systems w ill be conducted at Crane, IN and Seal Beach, CA. Training for the UPLC/PDA/MS/MS system w ill be conducted at Crane, IN.
CLIN 0004 Training to be conducted on site: NAVAL SURFACE WARFARE CENTER CRANE DIV
M/F: BENJAMIN WILKINS (BLDG 2707), 300 HWY 361, CRANE, IN 47522-5001
CLIN 0005 Training to be conducted on site: NSWC CORONA DIVISION SEAL BEACH DET, TRAM LE (TRACY) CHEMISTRY LAB, LRC, MS41- WEAPONS SURVEILLANCE & TESTING LAB, 800 SEAL BEACH BLVD BLD 5, SEAL BEACH CA 90740-5607 CLIN 0006 Training to be conducted on site: NAVAL SURFACE WARFARE CENTER CRANE DIV
M/F: BENJAMIN WILKINS (BLDG 2707), 300 HWY 361, CRANE, IN 47522-5001
1. CONTRACT ID CODE PAGE OF PAGES
J 1 3
16A. NAME AND TITLE OF CONTRACTING OFFICER (Type or print)
16C. DATE SIGNED
BY 25-Mar-2020
16B. UNITED STATES OF AMERICA15C. DATE SIGNED15B. CONTRACTOR/OFFEROR
(Signature of Contracting Officer)(Signature of person authorized to sign)
8. NAME AND ADDRESS OF CONTRACTOR (No., Street , County, State and Zip Code) X N0016420Q0099
X 9B. DATED (SEE ITEM 11)
19-Mar-2020
10B. DATED (SEE ITEM 13)
9A. AMENDMENT OF SOLICITATION NO.
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
X The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offer is extended, X is not extended.
Offer must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended by one of the following methods:
(a) By completing Items 8 and 15, and returning 1 copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted;
or (c) By separate letter or telegram which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY RESULT IN
REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by telegram or letter, provided each telegram or letter makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
12. ACCOUNTING AND APPROPRIATION DATA (If required)
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS.
IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.
A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE
CONTRACT ORDER NO. IN ITEM 10A.
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES (such as changes in paying office, appropriation date, etc.) SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(B).
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER (Specify type of modification and authority)
E. IMPORTANT: Contractor is not, is required to sign this document and return copies to the issuing office.
14. DESCRIPTION OF AMENDMENT/MODIFICATION (Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
10A. MOD. OF CONTRACT/ORDER NO.
2. AMENDMENT/MODIFICATION NO. 5. PROJECT NO.(If applicable)
6. ISSUED BY
3. EFFECTIVE DATE
25-Mar-2020
CODE
NAVAL SURFACE WARFARE CENTER CRANE DIV
S. CRAIG (812-854-6416)
BLDG 3373
300 HWY 361
CRANE IN 47522-5001
N00164 7. ADMINISTERED BY (If other than item 6)
4. REQUISITION/PURCHASE REQ. NO.
CODE
See Item 6
FACILITY CODECODE
EMAIL:TEL:
N0016420Q0099
SECTION SF 30 BLOCK 14 CONTINUATION PAGE
SECTION 4.0 - CLAUSES & OTHER FULL TEXT INFORMATION
SUMMARY OF CHANGES
SECTION 1.0 - SCHEDULE/INSPECTION & ACCEPTANCE/DELIVERY
The following have been modified:
SALIENT CHARACTERISTICS
Salient Characteristics UPLC/PDA The UPLC system must be able to operate and maintain pressure at or above 1300 PSI.
The UPLC system must include an autosampler.
The autosampler must be able to uptake and inject sample volumes as small as 3uL.
The autosampler compartment must be temperature controlled.
The autosampler must be able to hold a minimum of 48, 2.00mL vials samples.
The Injection needle must have a method of self-cleaning to prevent sample contamination.
The column compartment must be temperature controlled.
All instruments and components must be new and from the original manufacturer.
The equipment purchase must come with a service contract for all main components, and maintenance support must be certified by the manufacturer.
The PDA must be able to measure absorbance from 200nm to 800nm.
The UPLC pumping system must be at minimum a binary system.
UPLC must be able to operate in the pH range from 2 to 12.
The UPLCs must be identical for both the UPLC/PDA and UPLC/PDA/MS/MS.
Manufacturer must be able ship UPLC/PDA to Naval Surface Warfare Center, Corona Division, Detachment Seal Beach (California) and UPLC/PDA/MS/MS to Naval Surface Warfare Center, Crane Division (Indiana).
Manufacture must offer maintenance support to both Naval Surface Warfare Center, Crane Division (Indiana) and Naval Surface Warfare Center, Corona Division, Detachment Seal Beach (California).
Multiple day training for at least four people must be offered by the manufacturer for both the UPLC/PDA and UPLC/PDA/MS/MS systems. Training for the UPLC/PDA systems will be conducted at Crane, IN and Seal Beach, CA. Training for the UPLC/PDA/MS/MS system will be conducted at Crane, IN.
Software must be compatible with Windows 10.
Manufacturer must be able to provide an onsite visit for method development.
Instrumentation must include all required software operate each system and component of the UPLC/PDA and UPLC/PDA/MS/MS.
UPLC autosampler must be able to perform auto-dilution.
UPLC column compartment must be temperature controlled.
UPLC system must have column usage tracking technology.
Manufacture must be able to support extended warranties after purchase.
Manufacture must be able to perform manufacture installation of all equipment at both Seal Beach, CA and
Crane, IN locations
Salient Characteristics UPLC/PDA/MS/MS Note: The UPLC/PDA/MS/MS must include all of the same characteristics as the UPLC/PDA above and additionally it must meet the following criteria.
The Mass Analyzer must be able to perform with a high resolving power (>40,000 FWHM).
N0016420Q0099
The dynamic range of the mass spectrometer must be up to four orders of magnitude.
Mass Spectrometer must have a collision cell and be able to perform MS/MS product ion scanning.
All Tubing within the UPLC and MS must be free of fluoride containing compounds (i.e Teflon, viton, or
PEEK).
The Mass Spectrometer must be able to quantitatively determine concentrations of plasticizers down to 100 ppb with linearity coefficients greater than 0.995.
Mass Spectrometer must be able to qualitatively detect concentrations of fluoride containing (pre- and polyfluoroalkyl substances, PFAS) compounds down to parts per trillion.
Manufacturer must have user software that supports the understanding and identification of molecular structure of compounds in matrix systems based on the mass structures.
All software must be compatible with multiple UPLC/PDA/MS/MS libraries such as EPA and pre- and polyfluoroalkyl substances (PFAS) libraries.
Software must be able to perform calibration and determine sample concentration using both PDA and
MS/MS data.
Instrumentation package must come with vacuum pump and nitrogen generator to support MS/MS.
Mass spectrometer must have the capability to eliminate neutral ions Mass Spectrometer must come with a source that can perform both ESI and APCI simultaneously, without having to change the source.
Mass spectrometer must be able to readily accommodate different commercially available sources of ionization. Including DESI, DART, NanoESI, Unispray, ESI, APCI, and ESCi Mass spectrometer must come equipped with an Electro-Spray Ionization Source that is capable of ionizing a broad range of compounds.
Instrument must be able to perform Multiple Reaction Monitoring (MRM).
Instrument must be able to perform a full mass scan.
Instrument must be equipped with an internal calibration standard.
Instrument must have the ability to perform alternate scans between low and high collision energies (MSE).
Mass Spectrometer must be able to measure a mass range from m/z 20 to 100,000.
Mass sensitivity mode must be greater than 30,000 FWHM Manufacture must be able to support extended warranties after purchase.
Multiple day training for at least four people must be offered by the manufacturer for the
UPLC/PDA/MS/MS systems. Training for the UPLC/PDA/MS/MS system will be conducted at Crane, IN.
(End of Summary of Changes)