Minimum specification microNIR 042423.docx

DOCX document 20 KB Posted

Attached to
MicroNIR OnSite-W Federal contract opportunity
Solicitation number
12405B23Q0302
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery

View the file

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

USDA-ARS Sugarcane Production Research Unit 12990 US Hwy 441 N Canal Point, FL 33438

Minimum specifications

MicroNIR OnSite-W

Background and Scope This request is to purchase a handheld NIR instrument to conduct research.

Required Supporting Equipment

(1) Sample Shaker

(2) Tablet Probe

(3) Extended Warranty

(4) Technical Support

Applicable Documents and Standards Equipment should be shipped within standard shipment times after purchase is made.

Requirements Equipment must be a handheld, battery powered microNIR instrument that is able to connect wirelessly to a tablet PC or laptop. The NIR instrument should operate in the wavelength range 950-1650nm to match the wavelength of existing laboratory NIR instruments for comparison between the two types of instruments. Operating and data analysis software must be able to export collected data in CAMO .unsub, Grams .spc, and ASCII .csv file formats. Must have the option of attaching a probe to the handheld NIR instrument. Must operate between 20 and 40 C, must have signal to noise ratio of 25,000 with a typical measurement time of 0.25 to 0.5 seconds and a maximum dynamic range of 1000:1.

Testing and Verification Item must operate between 20 and 40 C, must have signal to noise ratio of 25,000 with a typical measurement time of 0.25 to 0.5 seconds and a maximum dynamic range of 1000:1.

Data Requirements Operating and data analysis software must be able to export collected data in CAMO .unsub, Grams .spc, and ASCII .csv file formats.

Packaging and Shipping Packaging and shipping requirements must comply with federal regulations for interstate shipments

Parameter
Specification
Illumination source
Two integrated vacuum tungsten lamps
Lamp life
>40,000 hr
Sample working distance
3 mm from window optimal, 0-15 mm
Dispersing element
VIAVI linear variable filter (LVF)
Detector
128 pixel InGaAs photodiode array
Wavelength range
950 – 1650 nm (10,526 – 6060 cm-1)
Pixel-to-pixel interval
6.2 nm for 950 – 1650 nm
Spectral bandwidth (FWHM)
<1.25% of center wavelength, 1% typical (e.g., @1000 nm, resolution is <12.5 nm)
Analog-to-digital convertor
16 bit
Dynamic range (Max)
1000:1
Measurement time (typical)
0.25 – 0.5 sec
Signal-to-noise ratio (typ.)
25,000
Integration time
10 ms typical, minimum 10 µsec
Connectivity
Bluetooth 4.1 (BLE) and USB 2.0
Weight
<250 g (<9 oz)
Size (Length x Diameter)
194 mm (6.64”) x 47 mm (1.85”)
Battery
Rechargeable Li-ion, not removable
Battery run time/charge
>10 hours with lamps on continuously
Battery cycle life
>75% capacity after 300 charge cycles
Battery charge time
3.5 hours at 1 A
Export data formats
CAMO .unsb, Grams .spc, ASCII .csv
Operating system
Microsoft Windows 10 Pro
Minimum screen resolution tablet/PC application
1920 x 1080/1366 x 768
Software compatibility
MicroNIR Pro 3.0 and above
Qualified PC
Microsoft Surface Pro 6 (tablet or PC)
Qualified Tablet PC
Dell Latitude 7212 Rugged Extreme
Operating temperature
−20 ~ 40°C (noncondensing)
Storage temperature
−20 ~ 50°C (noncondensing)
Dust and water resistance
IP65 and IP67
Shock and vibration
MIL-PRF-28800F Class 2

File details come from the government source that posted it. Updated .