Salient_Characteristics.pdf
PDF 27 KB Posted
- Attached to
- Eureka Water Probes (Or Equal) Federal contract opportunity
- Solicitation number
- G15PS00356
About this file
Salient Characteristics
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Sol_G15PS00356_Amd_000001.pdf | ||
| Sol_G15PS00356.pdf | ||
| Vendor_Instructions-Brand_Name_Or_Equal.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
The USGS Water Science Center in Northborough, MA requests the purchase of a multiprobe water quality sonde with low range depth, hand held display and cable. This equipment is intended to be used for manual and unattended monitoring of streams, rivers, lakes, and tidal wetlands. It must be capable of handling harsh environments. It is imperative that the equipment meets or exceeds the requirements below.
SALIENT CHARACTERISTICS
EUREKA WATER PROBES MANTA2, SUB 3.5, MLPS35-5P
Manta2:
1. Shall be easy to operate and be capable of handling harsh environments.
2. Shall not exceed 3-1/2 to 4 inches (88.9 to 101.6 mm) in diameter.
3. Shall weigh no more than 11 pounds (4990g).
4. Shall have an internal power supply using a maximum of six (6) 'C' size batteries.
5. Shall be capable of up to nine (9) sensors plus the temperature and pH ref.
6. Shall have an operating temperature of -5° to +50°C (23° to 122°F).
7. Shall be capable of communications using RS-232, SDI-12, USB, and Bluetooth.
8. Shall have sampling rate of 1Hz.
9. Shall have a data memory or 4MB;>1,000,000 logged readings.
10. Shall have a weighted sensor guard.
11. Shall have storage and calibration cups.
12. Shall have a low range depth sensor for up to 25 meters (82 feet).
13. Shall have parameter for temperature, 0.01 resolution, and ±0.1 accuracy.
14. Shall have parameter for specific conductance, 4 digits resolution, and ±0.3% of reading
±0.001 accuracy.
15. Shall have parameter for salinity, 0 to 70 PSS, 0.01 resolution, ±1% of reading ±0.1 PSS accuracy.
16. Shall have parameter for total dissolved solids (TDS), 0 to 65 g/1, 0.1 resolution, and ±5% of reading accuracy.
17. Shall have parameter for pH, 0 to 14 units, 0.01 resolution, ±0.1 within 10C of cal, ±0.2 otherwise accuracy.
18. Shall have parameter for optical dissolved oxygen (mg/1 and% saturation), 0.01 resolution, and ±1% of reading accuracy.
19. Shall have parameter for turbidity with wiper.
20. Shall have central wiping system.
21. Shall have a minimum of two (2) additional medium ports and two (2) small ports.
22. Shall have a flow cell of no more than 3.0 inches.
23. Shall have an underwater cable, 15 meters (49.2 feet) long.
24. Shall have quick connect/disconnect fittings on cable.
25. Shall include electronic manual.
26. Shall include control software.
27. Shall have a minimum of two (2) year warranty.
Amphibian2 Handheld Display:
28. Shall be made of durable plastic, shear-proof and shock-resistant; IP68 enclosure.
29. Shall be chemical resistant.
30. Shall be no more than 3.6" W x 7.25" L x 1.5'' D (91.4 x 184.2 x 38.1 mm).
31. Shall weigh no more than 1.3 pounds (590g) with battery.
32. Shall have a wide hand strap.
33. Shall have a minimum of 1.0 GHz processor.
34. Shall have Windows Handheld 6.5.3 operating system.
35. Shall have Microsoft Office Mobile 2010 software.
36. Shall have ports for COM1, RS-232, USB, 12V & 24VDC input.
37. Shall have Bluetooth and Wi-Fi connectivity options.
38. Shall have a minimum 4.3" (109 mm) viewing display, WVGS LCD TFT (800x480).
39. Shall have high visibility backlit LCD.
40. Shall be capable of both portrait and landscape orientation.
41. Shall have capacitive touch interface, chemically strengthened glass, scratch-resistant display screen.
42. Shall have a minimum 5MP resolution, autofocus, LED & video capture camera.
43. Shall have high-sensitivity GPS/GLONASS/SBAS receiver.
44. Shall have a minimum of 512MB RAM; 8GB flash storage.
45. Shall have an adjustable LED backlit keyboard with numeric keypad.
46. Shall have a minimum 3.7VDC rechargeable Li-Ion battery.
47. Shall include AC wall charger.
48. Shall have a tethered stylus.
49. Shall include data transfer cable.
50. Shall have a minimum of two (2) year warranty.
SPECIFICALLY MUST INCLUDE:
1 each, #MLP-S35-5P, MANTA2, SUB 3.5; includes weighted sensor guard, storage and calibration cups, temperature assembly, logging, electronic manual, Manta2 control software and standard full two year warranty
1 each, #SEN-A019, Low Range Depth Sensor (up to 25 meters)
1 each, #FlowCell-3.0, Flow Cell, 3.0"
1 each, #CBL-UW015, Cable, 15 meter, underwater
1 each, #DSP-AP2-MEAS, Amphibian2 Hand Held Display with capacitive touch interface, IP68 enclosure, Windows Embedded 6.5.3 OS, Manta2 GUI software, Microsoft Office Mobile
2010, Bluetooth, Wi-Fi, data transfer cable, 3.7V rechargeable battery, AC wall charger, 512MB
RAM/8GB flash storage, two year manufacturer's warranty
File details come from the government source that posted it. Updated .